Rocket Medical Plc recalls

2 recalls recorded since 2018, most recently on 7 November 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
7 November 2025
Reason
IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Distribution
Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, S
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0954-2026, Z-0955-2026, Z-0951-2026, Z-0952-2026, Z-0950-2026, Z-0953-2026

Terminated Class II Medical devices

Recall started
1 August 2019
Terminated
2 November 2021
Reason
A complaint was received regarding kinking of the device prior to use. If a device is kinked it cannot be used. In addition, the company became aware that the device had been used for pleural drainage, which is not included in the device indications.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0512-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.