Roche Molecular Systems, Inc. recalls

11 recalls recorded since 2018, most recently on 21 December 2022. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls11
Ongoing (FDA)4
Terminated or completed7
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
21 December 2022
Reason
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true diagnosis.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1105-2023, Z-1106-2023, Z-1104-2023

Terminated Class II Medical devices

Recall started
8 September 2022
Terminated
7 November 2024
Reason
Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.
Distribution
International distribution in the countries of Austria, Belgium, Czech Republic, France, Germany, Greece, Ireland, Italy, Lithuania, Poland, Portugal, Spain, Sweden, Switzerland, UK, Chile, Colombia, Hong Kong, Jamaica, Japan, Nicaragua, Oman, Pakistan, UAE.
Initiated
Voluntary: Firm initiated
Recall number
Z-0097-2023

Ongoing Class II Medical devices

Recall started
16 November 2021
Reason
Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.
Initiated
Voluntary: Firm initiated
Recall number
Z-0828-2022

Ongoing Class II Medical devices

Recall started
3 September 2021
Reason
False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, UK, Poland, Spain, Italy, Sweden, Greece, Denmark, Norway, Netherlands, France, Czech Republic, Belgium, Austria, Slovakia, Hungary, Switzerland, Lithuania, Slovenia. Serbia, North Macedonia, Myanmar, Lebanon, Russia, C
Initiated
Voluntary: Firm initiated
Recall number
Z-0083-2022

Terminated Class II Medical devices

Recall started
4 August 2021
Terminated
23 February 2024
Reason
Customers have reported an increased number of false positive SARS-CoV-2 results.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, CA, IL, IN, KS, ME, MI, MN, MO, NY, OR, TN, TX, WA, and WI. The countries of Canada, UK, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Netherlands, Norway, Poland, Slovakia, and Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-2498-2021

Ongoing Class II Medical devices

Recall started
25 March 2021
Reason
Invalid or false positive results for one or more targets (SARS-CoV-2/Influenza A/Influenza B) using the cobas SARS-CoV-2 & Influenza A/B assay for use under emergency use authorization on the cobas Liat analyzer, showed abnormal PCR growth curves
Distribution
Nationwide WorldWide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1414-2021, Z-1413-2021

Terminated Class II Medical devices

Recall started
23 November 2020
Terminated
22 May 2023
Reason
When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.
Distribution
Domestic distribution to Maryland and Michigan. International distribution to France, Greece, Italy, Poland, Turkey, Brazil, India, and Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-0877-2021

Terminated Class II Medical devices

Recall started
19 November 2020
Terminated
10 August 2023
Reason
Dirty Lens May Cause Invalid or False Positive Results
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Germany, Italy, Spain, United Kingdom, France, Austria, Turkey, Belgium, Poland, Portugal, Greece, Netherlands, Switzerland, Czech Republic, Romania, Sweden, Norway, Denmark, Hungary, Ireland, Slovakia, Croatia, Bulgari
Initiated
Voluntary: Firm initiated
Recall number
Z-0846-2021

Terminated Class II Medical devices

Recall started
3 June 2020
Terminated
4 October 2023
Reason
When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2551-2020

Terminated Class II Medical devices

Recall started
2 May 2018
Terminated
30 April 2019
Reason
Cross-contamination of samples has been reported when running the existing protocol on the system with the rack-based purification run.
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, GA, HI, IN, MD, ME, MI, MN, NC, NY, PA, VI, and WI. The products were distributed to the following foreign countries: Austria, Belgium, Brazil, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, France, Germany, Guatemala, Hong
Initiated
Voluntary: Firm initiated
Recall number
Z-1956-2018

Terminated Class II Medical devices

Recall started
27 October 2017
Terminated
27 April 2020
Reason
The cobas MRSA/SA tests may have decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others. The United States is not impacted, as the cobas MRSA/SA nucleic acid test for use on the cobas liat system is not available in the US.
Distribution
Worldwide distribution. Austria, Denmark, Germany, Singapore, Netherlands, Hong Kong, and Switzerland. US product is not affected.
Initiated
Voluntary: Firm initiated
Recall number
Z-1195-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.