Roche Diabetes Care, Inc. recalls

9 recalls recorded since 2018, most recently on 13 May 2024. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)2
Terminated or completed7
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
13 May 2024
Reason
The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.
Distribution
Nationwide and Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2174-2024

Terminated Class II Medical devices

Recall started
9 March 2022
Terminated
22 January 2025
Reason
Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.
Distribution
US Nationwide Distribution: AK, AZ, CA, CO, DC, FL, GA, IL, IN, MN, NC, NJ, NY, PA, RI, SC, TX, WI
Initiated
Voluntary: Firm initiated
Recall number
Z-1008-2022

Ongoing Class II Medical devices

Recall started
28 July 2021
Reason
Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2352-2021, Z-2351-2021, Z-2353-2021, Z-2354-2021

Terminated Class II Medical devices

Recall started
11 December 2019
Terminated
8 July 2022
Reason
Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a delay in treatment and/or therapy decisions
Distribution
Nationwide: Foreign: Austalia, India
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0997-2020, Z-0998-2020, Z-0999-2020

Terminated Class II Medical devices

Recall started
30 September 2019
Terminated
29 July 2020
Reason
Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
Distribution
US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0751-2020, Z-0752-2020, Z-0748-2020, Z-0747-2020, Z-0749-2020, Z-0750-2020

Terminated Class II Medical devices

Recall started
3 May 2019
Terminated
30 October 2020
Reason
Users with Android OS 8.0 and above may be unable to transfer values obtained on their meter to the App.
Distribution
Nationwide and worldwide usage of the Connect App.
Initiated
Voluntary: Firm initiated
Recall number
Z-2753-2019

Terminated Class II Medical devices

Recall started
7 May 2018
Terminated
16 January 2020
Reason
The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.
Distribution
Worldwide Distribution - US Nationwide distribution. Foreign distribution to Argentina Armenia Australia Austria Azerbaijan Bahamas Bahrain Barbados Belgium Brazil Bulgaria Canada Chile China Colombia Costa Rica Denmark Dominican Republic El Salvador Finland Georgia Germany Gha
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2186-2018, Z-2185-2018, Z-2187-2018

Terminated Class II Medical devices

Recall started
12 June 2017
Terminated
30 October 2020
Reason
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2765-2019, Z-2766-2019

Terminated Class II Medical devices

Recall started
8 June 2017
Terminated
27 July 2018
Reason
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0625-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.