Vancouver, Canada
Response Biomedical Corp. recalls
1 recall recorded since 2018, most recently on 15 March 2019. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 15 March 2019
- Terminated
- 9 June 2022
- Reason
- The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.
- Distribution
- Worldwide distribution: US Nationwide distributions in the states of IN, GA, OK, KS, AZ, SC, NE, NC, MI, FL, CA, MN, and AR. Countries of Canada, Italy, Malaysia, Kuwait, Saudi Arabia, UAE, Ukraine, Russia, Dominican Republic, France, Mexico, Colombia, Singapore, China, Serbia, Romania, and Phil
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1461-2020
- Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML".
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.