Los Angeles, CA, United States
RESPIRATORY THERAPEUTICS GROUP LLC recalls
1 recall recorded since 2018, most recently on 19 August 2019. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 19 August 2019
- Reason
- It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.
- Distribution
- US Nationwide distribution in the states of CA, NJ, NY. OUS: None
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1443-2020
- RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 -
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.