Ongoing Class I Medical devices

Recall started
25 June 2026
Reason
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Ar
Initiated
Voluntary: Firm initiated
Recall number
Z-2735-2026

Ongoing Class I Medical devices

Recall started
20 November 2023
Reason
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
Distribution
US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0538-2024, Z-0540-2024, Z-0541-2024, Z-0536-2024, Z-0537-2024, Z-0542-2024, Z-0539-2024

Ongoing Class I Medical devices

Recall started
13 September 2023
Reason
If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all
Distribution
Worldwide - US Nationwide distribution including in the states of CA, NJ, VA, IN, MI, WV, NY, WI, ND, IL, NC, FL, CO, TX, AL, LA, MD, OR, AZ, MN, WA, PA, SC, MA, ID, GA, SD, TN, NH, OH, MS, CT, IA, KY, UT, NE, MO, NV, AR, WY, RI, KS, VT, ME, OK, PR, MO, DC and the countries of Taiwan, Israel, Spain,
Initiated
Voluntary: Firm initiated
Recall number
Z-0111-2024

Completed Class I Medical devices

Recall started
12 December 2019
Reason
Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch.
Distribution
U.S.: PA, MA, VA, TN, CT, IL, GA, FL, MT
Initiated
Voluntary: Firm initiated
Recall number
Z-0915-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.