Recall Record

Farnborough, United Kingdom

Remote Diagnostic Technologies Ltd. recalls

12 recalls recorded since 2018, most recently on 26 November 2025. 10 are listed as ongoing. Status and classification are as published by the FDA.

Recalls12
Ongoing (FDA)10
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
26 November 2025
Reason
Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG
Initiated
Voluntary: Firm initiated
Recall number
Z-1027-2026

Ongoing Class II Medical devices

Recall started
30 May 2025
Reason
Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.
Distribution
Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.
Initiated
Voluntary: Firm initiated
Recall number
Z-2175-2025

Ongoing Class II Medical devices

Recall started
16 April 2025
Reason
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES
Initiated
Voluntary: Firm initiated
Recall number
Z-1666-2025

Ongoing Class II Medical devices

Recall started
15 April 2025
Reason
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE,
Initiated
Voluntary: Firm initiated
Recall number
Z-1729-2025

Ongoing Class III Medical devices

Recall started
28 October 2024
Reason
Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.
Distribution
U.S. Nationwide distribution in the states of AL, CA, FL, MO, NY, OH, TX and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0557-2025

Ongoing Class II Medical devices

Recall started
9 May 2024
Reason
Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.
Distribution
US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.
Initiated
Voluntary: Firm initiated
Recall number
Z-2028-2024

Ongoing Class II Medical devices

Recall started
24 August 2023
Reason
Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.
Distribution
US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE,
Initiated
Voluntary: Firm initiated
Recall number
Z-2625-2023

Ongoing Class II Medical devices

Recall started
8 May 2023
Reason
Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".
Distribution
US Nationwide distribution in the states of UT, CO, CA, NE, IL, TX, WV, KS, GA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1825-2023

Ongoing Class II Medical devices

Recall started
23 February 2023
Reason
During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.
Distribution
Worldwide Distribution: U.S.(nationwide): AL, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and O.U.S.(foreign): United Arab Emirates, Afghanistan, Austria, Australia, Belgium, Brunei
Initiated
Voluntary: Firm initiated
Recall number
Z-1321-2023

Terminated Class II Medical devices

Recall started
28 January 2022
Terminated
25 July 2024
Reason
Distribution of Defibrillators that are not approved or cleared for distribution US Market.
Distribution
US Nationwide distribution in the states of IL and KS.
Initiated
Voluntary: Firm initiated
Recall number
Z-0861-2022

Ongoing Class II Medical devices

Recall started
25 May 2021
Reason
The Tempus Pro (Trizeps 7 only) when used in combination with a specified laryngoscope device can cause two error types when subsequently unplugged.
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, TX, PA, MN, UT, MT, CA, WI, VA, NY, TN, KS, MD and the countries of Australia, Austria, Czech Republic, France, Germany, Hong Kong, Israel, Italy, Netherlands, Thailand, Abu Dhabi, UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-1997-2021

Terminated Class II Medical devices

Recall started
17 November 2020
Terminated
12 December 2023
Reason
A software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may put the ECG calibration out of calibration. This software error may result in a delay in therapy, as the clinician is not provided an ECG waveform on which a shockable rhythm decision can be made. The device is still able to deliver a manual defibrillation shock.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0564-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.