Remel Inc recalls

22 recalls recorded since 2018, most recently on 3 June 2026. 12 are listed as ongoing. Status and classification are as published by the FDA.

Recalls22
Ongoing (FDA)12
Terminated or completed10
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
3 June 2026
Reason
Reduced or no recovery of some strains of Neisseria gonorrhoeae.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Singapore, and the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2623-2026

Ongoing Class II Medical devices

Recall started
24 March 2026
Reason
Products may report incorrect AST results during quality control. Health consequences may include delayed response or the need to change to another antifungal agent if laboratory ignores Quality Control test result.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Vietnam, France, Argentina, Qatar, New Zealand, Uruguay, Turkey, Netherlands, Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-1806-2026

Ongoing Class II Medical devices

Recall started
5 February 2026
Reason
Customer complaints report low to no recovery of Campylobacter Jejuni ATCC 33291 on identified lot of medium.
Distribution
US Nationwide distribution in the states of AZ, CO, IL, NJ, OH, TN, WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1474-2026

Ongoing Class II Medical devices

Recall started
12 November 2025
Reason
Products may contain contamination, which may result in a darker or brown media color.
Distribution
Worldwide - US Nationwide distribution in the states of AK, CA, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, NE, NY, OH, OK, PA, RI, SD, TN, VA, WA, WI and the countries of Brazil, Canada, China, Cyprus, Finland, Great Britain, Ireland, Moldova, Norway, Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-0585-2026

Ongoing Class II Medical devices

Recall started
11 July 2025
Reason
The products may contain surface and subsurface contamination of Listeria monocytogenes.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2223-2025, Z-2226-2025, Z-2224-2025, Z-2225-2025

Ongoing Class II Medical devices

Recall started
1 May 2025
Reason
Product may have an off color affecting perfomance
Distribution
Worldwide - US Nationwide distribution in the states of AL, GA, KY, VA and the countries of Canada, UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-1947-2025

Ongoing Class II Medical devices

Recall started
13 February 2025
Reason
Product may contain surface and subsurface contamination of Listeria monocytogenes which is typically visible immediately upon removal from the packaging, appearing on the blood side of the biplate.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1201-2025, Z-1200-2025

Ongoing Class II Medical devices

Recall started
12 February 2025
Reason
Power supply unit for Zebra Printers used with the device can potentially overheat or cause a fire hazard.
Distribution
Domestic distribution to AL, AZ, CA, FL, GA, IN, KS, KY, MD, MI, NE, NY, OH, PA, TN, TX, WI. International distribution to Canada and the Bahamas.
Initiated
Voluntary: Firm initiated
Recall number
Z-1395-2025

Ongoing Class II Medical devices

Recall started
20 September 2024
Reason
A color change of the solution within the dropper can occur from orange-yellow to light yellow resulting in out-of-specification fluorescent stain results.
Distribution
Distribution was made to IL, IN, MN, NE, SD, TX, VA, and WA. There was government distribution but no military distribution. Foreign distribution was made to the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0266-2025

Ongoing Class II Medical devices

Recall started
14 August 2023
Reason
On lot of Todd Hewitt Broth w/CNA (LIM) may exhibit no recovery of target organism during customer QC testing or use with patient samples
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2531-2023

Ongoing Class II Medical devices

Recall started
31 May 2023
Reason
The test medium may not perform as intended.
Distribution
Distribution was made nationwide, including Puerto Rico. There was no foreign or military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2038-2023

Terminated Class I Medical devices

Recall started
20 October 2022
Terminated
30 September 2024
Reason
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0334-2023, Z-0329-2023, Z-0330-2023, Z-0328-2023, Z-0331-2023, Z-0333-2023, Z-0332-2023, Z-0335-2023, Z-0327-2023

Terminated Class II Medical devices

Recall started
10 January 2022
Terminated
24 July 2026
Reason
There are lower than expected MICs for some gram negative species.
Distribution
Distribution throughout the United States.
Initiated
Voluntary: Firm initiated
Recall number
Z-0722-2022

Terminated Class II Medical devices

Recall started
18 June 2021
Terminated
4 August 2023
Reason
Out of Specification
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, District of Columbia, Hawaii and Puerto Rico. The countries of BERMUD
Initiated
Voluntary: Firm initiated
Recall number
Z-2142-2021

Terminated Class II Medical devices

Recall started
22 April 2021
Terminated
8 August 2023
Reason
The media allows growth of E. faecalis ATCC 29212 with blackening of the media.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IL, IN, MA, MI, MN, MO, NC, ND, OH, OR, PA, SD, TN, TX, and VA. There was government distribution and no military or foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1561-2021

Ongoing Class II Medical devices

Recall started
10 March 2021
Reason
When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1425-2021, Z-1424-2021

Terminated Class II Medical devices

Recall started
10 December 2020
Terminated
6 July 2023
Reason
The blood agar side of the bi-plate may not grow some patient isolates or QC organisms as intended.
Distribution
Distribution in United States - AL, FL, GA, LA MS, NC, SC, TN
Initiated
Voluntary: Firm initiated
Recall number
Z-0769-2021

Terminated Class II Medical devices

Recall started
17 April 2019
Terminated
21 August 2020
Reason
Potential for products to fail performance testing per the Instructions for Use (IFU) starting at 57 days after date of manufacturing. Expiration date is shortened from 7 months to 56 days.
Distribution
Worldwide Distribution -US Nationwide distribution to AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, VA, VT, WA, WI, WV, WY, Puerto Rico. International distribution to Singapore
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1377-2019, Z-1378-2019, Z-1380-2019, Z-1379-2019

Terminated Class II Medical devices

Recall started
17 April 2019
Terminated
10 July 2020
Reason
Potential for products to fail performance testing per the Instructions for Use (IFU) starting at 106 days after date of manufacturing. Expiration date is shortened from 9 months to 105 days.
Distribution
Worldwide Distribution -US Nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico. International distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1409-2019, Z-1404-2019, Z-1405-2019, Z-1407-2019, Z-1402-2019, Z-1408-2019, Z-1403-2019, Z-1400-2019, Z-1406-2019, Z-1401-2019

Terminated Class II Medical devices

Recall started
4 January 2019
Terminated
17 August 2020
Reason
Potential for out of range microbial results
Distribution
...
Initiated
Voluntary: Firm initiated
Recall number
Z-1939-2020

Terminated Class II Medical devices

Recall started
13 August 2018
Terminated
18 June 2019
Reason
Products have intermittently failed performance testing during stability studies with various anaerobic microorganisms.
Distribution
US, Canada, Singapore
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3019-2018, Z-3018-2018, Z-3017-2018

Terminated Class II Medical devices

Recall started
8 December 2017
Terminated
22 July 2020
Reason
The serum may fail to agglutinate within the specified minimum reaction time.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-0290-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.