4 recalls recorded since 2018, most recently on 9 February 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)3
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
9 February 2026
Reason
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Distribution
US: CA, TX, PA, CT, NJ, OR, LA, OH
Initiated
Voluntary: Firm initiated
Recall number
Z-1656-2026

Ongoing Class II Medical devices

Recall started
22 July 2025
Reason
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
Distribution
U.S. Nationwide distribution in the states of CA, CT, LA, NJ, OR, PA, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2486-2025

Ongoing Class II Medical devices

Recall started
16 April 2024
Reason
A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.
Distribution
U.S.: CA, CT, NJ, OR, PA and TX No O.U.S.
Initiated
Voluntary: Firm initiated
Recall number
Z-1933-2024

Terminated Class II Medical devices

Recall started
24 November 2021
Terminated
8 July 2024
Reason
Due to dose discrepancy when delivering a plan to a patient in a Non-HFS (Head First Supine) orientation specifically in Feet First Supine (FFS).
Distribution
U.S. Nationwide distribution in the states of CA and TX
Initiated
Voluntary: Firm initiated
Recall number
Z-0470-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.