RAYSEARCH LABORATORIES AB recalls
37 recalls recorded since 2018, most recently on 20 May 2026. 16 are listed as ongoing. Status and classification are as published by the FDA.
Recalls37
Ongoing (FDA)16
Terminated or completed21
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 20 May 2026
- Reason
- Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2646-2026, Z-2647-2026
- RayCare 2024A SP1. Oncology Information Systems.
- RayCare 2024A SP4. Oncology Information Systems.
Ongoing Class II Medical devices
- Recall started
- 28 November 2025
- Reason
- Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
- Distribution
- Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1110-2026, Z-1106-2026, Z-1108-2026, Z-1109-2026, Z-1104-2026, Z-1105-2026, Z-1103-2026, Z-1107-2026, Z-1102-2026
- Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System
- Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Product Description: Radiation Therapy T
- Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning Syst
- Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 Software Version: RayStation 2024B Product Description: Radiation Therapy Treatment Planning System Component: No
- RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therap
- Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation
- RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System
- Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
- and 1 more product lines
Ongoing Class II Medical devices
- Recall started
- 8 August 2025
- Reason
- DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2492-2025, Z-2488-2025, Z-2491-2025, Z-2489-2025, Z-2494-2025, Z-2490-2025, Z-2493-2025, Z-2487-2025, Z-2495-2025
- RayStation with the following product descriptions: 1. RayStation 11B, Software Version: 12.0.0.932. 2. RayStation 11B Service Pack 1, Software Version: 12.1.0.1221. 3. RayStation 11B Service Pack 2, Software V
- RayStation with the following product descriptions: 1. RayStation 8B, Software Version: 8.1.0.47. 2. RayStation 8B Service Pack 1, Software Version: 8.1.1.8. 3. RayStation 8B Service Pack 2, Software Version: 8
- RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Ver
- RayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2. RayStation 9B Service Pack 1, Software Version: 9.2.0.483. 3. RayStation 10A, Software Version: 10.0.0.1154.
- RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1, Software Version: 6.1.1.2. 2. RayStation 6 Service Pack 2, Software Version: 6.2.0.7. 3. RayStation 6 Service Pack 3, Software
- RayStation with the following product descriptions: 1. RayStation 10B, Software Version: 10.1.0.613. 2. RayStation 10B Service Pack 1, Software Version: 10.1.1.54. Product Description: Radiation Therapy Treat
- RayStation with the following product descriptions: 1. RayStation 12A, Software Version: 13.0.0.1547. 2. RayStation 12A Service Pack 1, Software Version: 13.1.0.144. 3. RayStation 12A Service Pack 2, Software V
- RayStation with the following product descriptions: 1. RayStation 7, Software Version: 7.0.0.19. 2. RayStation 8A, Software Version: 8.0.0.61. 3. RayStation 8A Service Pack 1, Software Version: 8.0.1.10. P
- and 1 more product lines
Ongoing Class II Medical devices
- Recall started
- 16 January 2025
- Reason
- Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1189-2025, Z-1190-2025, Z-1186-2025, Z-1188-2025, Z-1195-2025, Z-1185-2025, Z-1187-2025, Z-1192-2025, Z-1191-2025, Z-1194-2025, Z-1193-2025
- Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3
- Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Version: RayStation 11B and 11B Service Pack
- Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113 and 9.0.1.142 Software Version: RayStation 8B, 8B Service Pack 1 and 2, 9A a
- Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1
- Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4 and 4.7.6.7 Software Version: RayStation 4.5.1, 4.7.2, 4.7.3, 4.7.4,
- Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1
- Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52 and 10.0.2.10 Software Version: RayStation 9B, 9B Service Pack 1, 1
- Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.1.1.2, 6.2.0.7, 6.3.0.6 Software Version: RayStation 6 Service Pack 1, 2 and 3
- and 3 more product lines
Ongoing Class II Medical devices
- Recall started
- 28 March 2024
- Reason
- Potential for reported SSD to be too high.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zeal
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1750-2024, Z-1743-2024, Z-1744-2024, Z-1748-2024, Z-1742-2024, Z-1752-2024, Z-1741-2024, Z-1746-2024, Z-1747-2024, Z-1749-2024, Z-1751-2024, Z-1745-2024
- RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
- RayStation 8.1.0.47, 8.1.1.8 and 8.1.2.5. Radiation Therapy Treatment Planning System.
- RayStation 9.0.0.113 and 9.0.1.142. Radiation Therapy Treatment Planning System.
- RayStation 11.0.0.951, 11.0.1.29, 11.0.3.116 & 11.0.4.15. Radiation Therapy Treatment Planning System.
- RayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.
- RayStation 15.0.0.430. Radiation Therapy Treatment Planning System.
- RayStation 7.0.0.19. Radiation Therapy Treatment Planning System.
- RayStation 10.0.0.1154, 10.0.1.52 & 10.0.2.10. Radiation Therapy Treatment Planning System.
- and 4 more product lines
Ongoing Class II Medical devices
- Recall started
- 23 February 2024
- Reason
- A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hong Kong S.A.R., Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1482-2024
- RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
Ongoing Class II Medical devices
- Recall started
- 12 September 2023
- Reason
- An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.
- Distribution
- MD, TN
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0206-2024
- RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (buil
Ongoing Class II Medical devices
- Recall started
- 17 July 2023
- Reason
- An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
- Distribution
- US Nationwide distribution in the state of TN.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2464-2023
- RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for
Ongoing Class II Medical devices
- Recall started
- 17 February 2023
- Reason
- It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
- Distribution
- US Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1281-2023
- Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4
Terminated Class II Medical devices
- Recall started
- 10 June 2022
- Terminated
- 11 June 2024
- Reason
- This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
- Distribution
- US Nationwide distribution in the states of NJ and TX.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1299-2022
Terminated Class II Medical devices
- Recall started
- 1 June 2022
- Terminated
- 4 October 2023
- Reason
- An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
- Distribution
- Worldwide distribution - US Nationwide distribution in the state of TN and the countries of Belgium and Switzerland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1403-2022
- RayStation 9B SP1. For radiation treatment planning.
Ongoing Class II Medical devices
- Recall started
- 30 December 2021
- Reason
- If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running
- Distribution
- US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0732-2022
- RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8
Ongoing Class II Medical devices
- Recall started
- 22 September 2021
- Reason
- An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be both under- and overestimated
- Distribution
- US Nationwide.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0849-2022
- RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.
Terminated Class II Medical devices
- Recall started
- 9 September 2021
- Terminated
- 16 April 2024
- Reason
- If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS, NC, OK, OR, SC, TN, WA, WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0848-2022
- TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483
Terminated Class II Medical devices
- Recall started
- 8 June 2021
- Terminated
- 20 March 2024
- Reason
- Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.
- Distribution
- US distribution to TN; and Belgium
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2138-2021
- RayStation (Radiation Treatment Planning System) : RayStation 9A, 9B, 9B SP1, 10A, 10A SP1, 10B and 11A, in combination with RayCare Model Numbers/ UDI: RayStation 9.0 (9.0.0.113) 07350002010174; RayStation 9.1 (9
Terminated Class II Medical devices
- Recall started
- 21 April 2021
- Terminated
- 27 July 2023
- Reason
- Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AZ, CA, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, Colombia
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1838-2021
- RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 ,
Terminated Class II Medical devices
- Recall started
- 7 April 2021
- Terminated
- 2 June 2023
- Reason
- When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.
- Distribution
- Domestic: TN
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1598-2021
- RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs
Ongoing Class II Medical devices
- Recall started
- 25 March 2021
- Reason
- Flags edited in RayCare after a patient has been checked in may not always be updated in RayTreat.
- Distribution
- US Nationwide in the state of TN.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1427-2021
- RayStation Model Number: (9A)9.0.0.113, 9B (9.1.0.933), 9B Service pack (9.2.0.483), 10A (10.0.0.1154), 10A Service pack (10.0.1.52), 10B (10.1.0.613)
Ongoing Class II Medical devices
- Recall started
- 3 March 2021
- Reason
- Patient related messages created in RayCare 3B, RayCare 4A will be lost when upgrading from RayCare 2 to RayCare 3B or later
- Distribution
- US Nationwide distribution in the state of TN.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1281-2021
- RayCare 3B (3.1.0.60341) - Product Usage: Intended Use Oncology information system that supports healthcare professionals in delivering cancer care.
Terminated Class II Medical devices
- Recall started
- 22 February 2021
- Terminated
- 1 May 2023
- Reason
- For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.
- Distribution
- US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1597-2021
- RayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 1, 2, 7, 8A, 8B, 9A, 10A, 10B including some of the Service Packs. Standalone software treatment planning system
Terminated Class II Medical devices
- Recall started
- 11 November 2020
- Terminated
- 9 May 2023
- Reason
- Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.
- Distribution
- US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0589-2021
- RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7
Terminated Class II Medical devices
- Recall started
- 8 October 2020
- Terminated
- 30 March 2023
- Reason
- If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.
- Distribution
- US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0478-2021
- RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy a
Ongoing Class II Medical devices
- Recall started
- 29 July 2020
- Reason
- When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image selected. When starting the task, an error message might be displayed stating: Review files not available.
- Distribution
- US Nationwide distribution including in the state of Tennessee.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2899-2020
- RayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care.
Ongoing Class II Medical devices
- Recall started
- 26 May 2020
- Reason
- It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.
- Distribution
- Distributed to one consignee in Nashville. Foreign distribution to one consignee in Belgium.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2426-2020
- RayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2
Terminated Class II Medical devices
- Recall started
- 18 May 2020
- Terminated
- 6 August 2021
- Reason
- Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
- Distribution
- US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2342-2020, Z-2345-2020, Z-2340-2020, Z-2343-2020, Z-2349-2020, Z-2346-2020, Z-2351-2020, Z-2344-2020, Z-2353-2020, Z-2348-2020, Z-2339-2020, Z-2341-2020 and 3 more
- RayStation 6 Service Pack 2, UDI # 07350002010075 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 8A Service Pack 1, UDI # 07350002010136 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 5 Service Pack 3, UDI # 07350002010020 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 6 Service Pack 3, UDI # 07350002010242 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 9A, UDI # 07350002010174 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 8B, UDI # 07350002010129 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 9B Service Pack 1, UDI # 07350002010297 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- RayStation 8A, UDI # 07350002010112 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- and 7 more product lines
Ongoing Class II Medical devices
- Recall started
- 4 March 2020
- Reason
- Due to physical limitations of the treatment machine, the maximum field opening for Siemens Virtual Wedge fields recommended by the LINAC vendor will be more restricted in the wedged direction than the maximum field opening without a wedge. Delivering a larger field opening with a virtual wedge will not result in a wedged field shape over the entire field. Instead, a high dose plateau region will be the result at the toe side of the field beyond the recommended field size.
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, and WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1598-2020
- RayStation standalone software treatment planning system, Model Numbers 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 7.0, 8.0, 8.1, 9.0, 9.1 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. B
Terminated Class II Medical devices
- Recall started
- 20 February 2020
- Terminated
- 5 November 2024
- Reason
- There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- Distribution
- The products were distributed to the following US states: TN
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1466-2020, Z-1463-2020, Z-1465-2020, Z-1468-2020, Z-1464-2020, Z-1462-2020, Z-1467-2020
- RayStation/RayPlan 8B Service Pack 2, Build Number 8.1.2.5, stand-alone software treatment planning system
- RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system
- RayStation/RayPlan 8B Service Pack 1, Build Number 8.1.1.8, stand-alone software treatment planning system
- RayStation/RayPlan 9B, Build Number 9.1.0.933, stand-alone software treatment planning system
- RayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone software treatment planning system
- RayStation/RayPlan 8A, Build Number 8.0.0.61, stand-alone software treatment planning system
- RayStation/RayPlan 9A, Build Number 9.0.0.113, stand-alone software treatment planning system
Terminated Class II Medical devices
- Recall started
- 19 December 2019
- Terminated
- 24 November 2020
- Reason
- Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf behavior. There is an interoperability issue with Elekta regarding setting of guard leaves. iii)SSD, when intended as source-to-surface distance, it sometimes gives source-to-skin distance.To the best of our knowledge, these issues have not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect dose calculations during treatment planning.
- Distribution
- US Nationwide distribution in states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, WI
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0866-2020
- RayStation stand-alone software treatment planning system. Models 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0.
Terminated Class II Medical devices
- Recall started
- 18 November 2019
- Terminated
- 10 November 2020
- Reason
- Editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. If the bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported.
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, and WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0736-2020
- RayStation stand-alone software treatment planning system, Model Numbers 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0 - Product Usage: RayStation is a software system designed for treatment planning an
Terminated Class II Medical devices
- Recall started
- 12 November 2018
- Terminated
- 17 April 2020
- Reason
- Isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system.
- Distribution
- AZ CA CO CT DE FL GA HI ID IL LA MA MD ME MI MO MS NC NJ NY OH OR PA RI TN TX VA VT WA WI
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0679-2019
- RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A
Terminated Class II Medical devices
- Recall started
- 21 September 2018
- Terminated
- 16 June 2020
- Reason
- Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
- Distribution
- Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0294-2019
- RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system.
Terminated Class II Medical devices
- Recall started
- 29 June 2018
- Terminated
- 16 July 2020
- Reason
- If the beam model has a highly asymmetric primary source, it is not correctly taken into account in the calculation of DMLC fields when the collimator is rotated. This could lead to potentially significant overdosage at delivery. The user must be aware of the issue to avoid incorrect dose calculations during treatment planning.
- Distribution
- US Nationwide Distribution to accounts in AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, and WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2497-2018
- RayStation stand-alone software treatment planning system, RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) and RayStation 8A (RayPlan 8A) P
Terminated Class II Medical devices
- Recall started
- 6 June 2018
- Terminated
- 4 April 2019
- Reason
- The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must be aware to avoid incorrect dose calculations during treatment planning.
- Distribution
- US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2554-2018
- RayStation Radiation Therapy Treatment Planning System, Model nos. 2.5, 3.5, 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 7.0 Product Usage RayStation is a software system designed for treatment planning and analysis
Terminated Class II Medical devices
- Recall started
- 28 March 2018
- Terminated
- 27 May 2020
- Reason
- The dose calculation accuracy may in some situations be less than expected. The user must be aware in order to avoid incorrect dose calculations during treatment planning.
- Distribution
- US Nationwide Distribution in the states of AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, ID, IL, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NY, OH, OR, PA, TN, TX, UT, VA, VT, WA, WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2290-2018
- RayStation Radiation Therapy Treatment Planning System; 6.0, 6.1, 6.2, 7.0 Product Usage: Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans
Terminated Class II Medical devices
- Recall started
- 26 January 2018
- Terminated
- 13 January 2020
- Reason
- Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient.
- Distribution
- Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1275-2018
- Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19 UDI: 0735000201006820171130
Terminated Class II Medical devices
- Recall started
- 20 November 2017
- Terminated
- 8 August 2023
- Reason
- When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong
- Distribution
- US Nationwide Distribution
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1930-2019
- RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6) . Intended Use: software system designed for treatment planning and ana
Terminated Class II Medical devices
- Recall started
- 17 July 2017
- Terminated
- 29 August 2018
- Reason
- For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
- Distribution
- US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0380-2018
- RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expe