Radiometer America Inc recalls

4 recalls recorded since 2018, most recently on 10 July 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)0
Terminated or completed4
Class I (FDA)0

Terminated Class II Medical devices

Recall started
10 July 2019
Terminated
13 June 2022
Reason
Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.
Distribution
US: AB, AL, AR, BC, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TX, VA, VT, WA, WI, WV OUS: Canada, Netherland, China, Czech Republic, Denmark, Germany, Spain, France, Hugary, , indiat Japan, South Korea, United Kingdom, Austral
Initiated
Voluntary: Firm initiated
Recall number
Z-2320-2019

Terminated Class II Medical devices

Recall started
3 July 2018
Terminated
30 August 2023
Reason
The analyzer software may cause a mis-match of patient demographics and test results when a certain sequence of events occur.
Distribution
Distribution was nationwide. There was also military/government distribution. Foreign distribution was made to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0682-2019

Terminated Class II Medical devices

Recall started
2 July 2018
Terminated
2 April 2024
Reason
From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.
Distribution
US: AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, PA, TN, TX, VA, WA, WI, and WV OUS: Canada, Croatia, Finland and Russia
Initiated
Voluntary: Firm initiated
Recall number
Z-0801-2019

Terminated Class II Medical devices

Recall started
25 May 2018
Terminated
6 November 2020
Reason
The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again
Distribution
US Distribution and Internationally to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0010-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.