Recall Record

San Diego, CA, United States

QUIDEL CARDIOVASCULAR INC recalls

2 recalls recorded since 2018, most recently on 25 May 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
25 May 2023
Reason
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
Distribution
US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL
Initiated
Voluntary: Firm initiated
Recall number
Z-1974-2023

Terminated Class II Medical devices

Recall started
9 April 2019
Terminated
14 January 2021
Reason
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
Distribution
US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL
Initiated
Voluntary: Firm initiated
Recall number
Z-1552-2020