Qiagen Sciences LLC recalls

19 recalls recorded since 2018, most recently on 12 March 2025. 8 are listed as ongoing. Status and classification are as published by the FDA.

Recalls19
Ongoing (FDA)8
Terminated or completed11
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
12 March 2025
Reason
Identified faulty cartridges. If such a faulty cartridge is used for sample testing, false test results could occur.
Distribution
US Distribution: AL, CA, DC, FL, IN, KS, KY, MA, ME, MN, NY, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1504-2025

Ongoing Class II Medical devices

Recall started
3 June 2024
Reason
Heater shaker module does not perform heating if the temperature is set below 40¿C for static incubation steps when running IVD protocol QIAamp¿ DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.
Distribution
Nationwide. Foreign: AD, AT, AU, BE, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB, ID, IN, IT, JO, KR, KZ, LU, ML, MY, NL, NO, PH, PL, PY, SA, SE, SG, SK, SO, VN J, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI
Initiated
Voluntary: Firm initiated
Recall number
Z-2353-2024

Ongoing Class II Medical devices

Recall started
4 March 2024
Reason
Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost
Distribution
Nationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO, NL, NO, QA, SA, SE, SI, TH
Initiated
Voluntary: Firm initiated
Recall number
Z-1508-2024

Ongoing Class II Medical devices

Recall started
28 April 2023
Reason
Faulty cartridges in the lot could result in false test results.
Distribution
US Nationwide distribution in the states of CA, FL, IL, MD, MS, NY, OK, PA, TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-1683-2023

Ongoing Class II Medical devices

Recall started
20 July 2022
Reason
Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the patient being exposed to an inappropriate or suboptimal anti-cancer treatment and/or unnecessarily experiencing side effects
Distribution
IN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands, Malaysia, Spain, Thailand
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1652-2022, Z-1651-2022, Z-1650-2022

Ongoing Class II Medical devices

Recall started
25 February 2022
Reason
False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.
Distribution
AZ CA CT FL GA IL IN MA MN NC NY OK OR PA TN TX WV
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0951-2022, Z-0952-2022

Ongoing Class II Medical devices

Recall started
21 November 2021
Reason
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
Distribution
Domestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN, NV, NJ, NY, NC, OH, PA, SC, SD, TN, TX, VA, WA, WI. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1746-2022

Terminated Class II Medical devices

Recall started
8 November 2021
Terminated
29 March 2023
Reason
The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
Distribution
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0455-2022

Terminated Class II Medical devices

Recall started
15 January 2021
Terminated
11 July 2022
Reason
QIAGEN has become aware of the potential for false positive results to occur with some patient samples.
Distribution
Domestic: AZ, CA, DC, CO,FL,HI, IL, KY, MI, MO, NC, NJ, NM, NY, OH, RI, TX, WA, WY
Initiated
Voluntary: Firm initiated
Recall number
Z-0941-2021

Terminated Class II Medical devices

Recall started
11 December 2020
Terminated
22 April 2022
Reason
Kit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instructions For Use.
Distribution
CA, IN, MI, MN, NC, NJ, NM, OH, TX
Initiated
Voluntary: Firm initiated
Recall number
Z-0937-2021

Terminated Class II Medical devices

Recall started
15 July 2020
Terminated
22 March 2021
Reason
There is an increased rate of potential false positive results for the Influenza A (no subtype) target, which may lead to 1) inappropriate antiviral use, 2) erroneous or delayed differential diagnosis and missing of an alternative diagnosis.
Distribution
US Nationwide distribution including in the states of AK, AL, CA, DC, DE, FL, GA, IL, IN, KY, MA, MD, MN, NC, NJ, NY, OH, OR, PA, TX, WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2938-2020

Terminated Class II Medical devices

Recall started
8 April 2020
Terminated
3 June 2021
Reason
There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-2290-2020

Terminated Class II Medical devices

Recall started
21 February 2020
Terminated
5 October 2020
Reason
An image on the associated polybag contains an incorrect rinse time of 30 (thirty seconds). The correct rinse time is 1 (one) minute.
Distribution
Nationwide domestic distribution. Foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1636-2020

Terminated Class II Medical devices

Recall started
28 January 2020
Terminated
26 September 2024
Reason
When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment
Distribution
Worldwide distribution - US Nationwide distribution and countries Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cameroon, Canada, Chile, China, Colombia, Congo, Costa Ric
Initiated
Voluntary: Firm initiated
Recall number
Z-1476-2020

Ongoing Class II Medical devices

Recall started
2 July 2019
Reason
QIAsymphony software version 5.0.3 - Software issue incorreclty assigns the 2D bar code eluate ID to the sample within the results file. If 2D bar code eluate IDs are not checked against the sample ID prior to downstream application the incorrect tube may be selected and it could have the potential to lead to delayed results or incorrect patient reporting
Distribution
Worldwide Distribution - US Nationwide Foreign: Algeria Australia Austria Azerbaijan Belgium Brazil Canada China Czech Republic Denmark Finland France Germany Greece India Israel Italy Japan Jordan Korea, Republic of Kuwait Latvia Mexico Netherlands Norway Pakistan Poland
Initiated
Voluntary: Firm initiated
Recall number
Z-2319-2019

Terminated Class I Medical devices

Recall started
5 June 2019
Terminated
28 August 2020
Reason
Kit lots contain a small percentage of damaged Filter-Tips that are leaking when being used with the QIAsymphony SP/AS Instruments (REF 9001297 and 9001301)
Distribution
US Nationwide distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2088-2019

Terminated Class II Medical devices

Recall started
6 May 2019
Terminated
23 April 2020
Reason
The deparaffinization solution contains a contaminant, which leads to reduced lysis efficiency of the affected lot, resulting in significantly reduced yields and less sensitive downstream applications.
Distribution
US nationwide distribution in the states of MA, RI, AZ, NY, NC, IN, TX. The products were distributed to the following foreign countries: Germany, Great Britain, Italy, France, Sweden, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1891-2019

Terminated Class II Medical devices

Recall started
23 July 2018
Terminated
5 June 2020
Reason
An issue with the power branching printed circuit board (PCB) on the instrument that has the potential to trigger a heater malfunction was discovered. If this failure were to occur, nucleic acid extractions could be compromised and extractions need to be repeated. This could cause a delay in obtaining results from downstream applications.
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, GA, IA, ID, IL, MA, MD, MI, MN, NE, NJ, NY, NV, OH, PA, PR, SD, VA, and WA. The products were distributed to the following foreign countries: Alger, Argentina, Australia, Austria, Canada, China, Dubai, Egypt, France, Germany,
Initiated
Voluntary: Firm initiated
Recall number
Z-0383-2019

Terminated Class I Medical devices

Recall started
18 May 2018
Terminated
2 September 2020
Reason
Potential for control line not appearing on the test strip
Distribution
Worldwide - US Nationwide distribution an in the country of , Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-2577-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.