Primus Corporation dba Trinity Biotech recalls

1 recall recorded since 2018, most recently on 5 October 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class III Medical devices

Recall started
5 October 2021
Terminated
9 August 2023
Reason
The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
Distribution
Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0349-2023, Z-0350-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.