Recall Record

Cortlandt Manor, NY, United States

Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc recalls

1 recall recorded since 2018, most recently on 30 April 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
30 April 2018
Terminated
17 April 2019
Reason
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Distribution
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3185-2018, Z-3186-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.