Philips North America Llc recalls

161 recalls recorded since 2018, most recently on 28 August 2026. 101 are listed as ongoing. Status and classification are as published by the FDA.

Recalls161
Ongoing (FDA)101
Terminated or completed60
Class I (FDA)4

Ongoing Class II Medical devices

Recall started
28 August 2026
Reason
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Distribution
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Gree
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3309-2026, Z-3313-2026, Z-3308-2026, Z-3318-2026, Z-3316-2026, Z-3312-2026, Z-3315-2026, Z-3307-2026, Z-3310-2026, Z-3311-2026, Z-3317-2026, Z-3314-2026

Ongoing Class II Medical devices

Recall started
21 August 2026
Reason
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3233-2026, Z-3239-2026, Z-3242-2026, Z-3241-2026, Z-3236-2026, Z-3234-2026, Z-3235-2026, Z-3238-2026, Z-3240-2026, Z-3237-2026, Z-3232-2026, Z-3243-2026

Ongoing Class II Medical devices

Recall started
18 August 2026
Reason
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Distribution
U.S. and U.S. Territories.
Initiated
FDA Mandated
Recall number
Z-3155-2026

Ongoing Class II Medical devices

Recall started
7 August 2026
Reason
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Distribution
US and US Territories.
Initiated
FDA Mandated
Recall numbers
Z-3043-2026, Z-3042-2026, Z-3041-2026, Z-3040-2026

Ongoing Class II Medical devices

Recall started
31 July 2026
Reason
Due to software malfunction, patient information system may freeze and become unresponsive.
Distribution
Worldwide - US Nationwide distribution including in the states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR and the cou
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3142-2026, Z-3141-2026

Ongoing Class II Medical devices

Recall started
1 July 2026
Reason
Device will intermittently provide inaccurate data related to native heart rate.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, France, Germany, Gibraltar, Hong Kong, Iceland, Ireland, Iraq, Italy, Luxembourg, Malta, Netherlands, New Caledonia, New Zealand, Norway, Pol
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2955-2026, Z-2956-2026

Ongoing Class II Medical devices

Recall started
5 June 2026
Reason
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-2579-2026

Ongoing Class II Medical devices

Recall started
26 May 2026
Reason
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
Distribution
Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2522-2026, Z-2523-2026

Ongoing Class II Medical devices

Recall started
1 May 2026
Reason
System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events.
Distribution
US distribution to California.
Initiated
Voluntary: Firm initiated
Recall number
Z-2354-2026

Ongoing Class II Medical devices

Recall started
14 April 2026
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1966-2026, Z-1968-2026, Z-1958-2026, Z-1955-2026, Z-1960-2026, Z-1951-2026, Z-1957-2026, Z-1963-2026, Z-1959-2026, Z-1952-2026, Z-1956-2026, Z-1964-2026 and 7 more

Ongoing Class II Medical devices

Recall started
7 January 2026
Reason
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egy
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1315-2026, Z-1317-2026, Z-1316-2026, Z-1318-2026

Ongoing Class II Medical devices

Recall started
22 December 2025
Reason
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzer
Initiated
Voluntary: Firm initiated
Recall number
Z-1156-2026

Ongoing Class II Medical devices

Recall started
12 December 2025
Reason
It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1033-2026

Ongoing Class II Medical devices

Recall started
3 December 2025
Reason
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, Unite
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1215-2026, Z-1209-2026, Z-1204-2026, Z-1207-2026, Z-1213-2026, Z-1208-2026, Z-1212-2026, Z-1205-2026, Z-1216-2026, Z-1210-2026, Z-1211-2026, Z-1214-2026 and 1 more

Ongoing Class II Medical devices

Recall started
31 October 2025
Reason
Potential issue where the IntelliVue monitors did not alarm.
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Botswana, Brazil, Brunei, Bulgaria, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Co
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0858-2026, Z-0852-2026, Z-0869-2026, Z-0867-2026, Z-0864-2026, Z-0870-2026, Z-0849-2026, Z-0854-2026, Z-0857-2026, Z-0861-2026, Z-0853-2026, Z-0859-2026 and 10 more

Ongoing Class II Medical devices

Recall started
29 October 2025
Reason
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Rep
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0882-2026, Z-0881-2026

Ongoing Class II Medical devices

Recall started
27 October 2025
Reason
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Distribution
U.S.
Initiated
FDA Mandated
Recall number
Z-0597-2026

Ongoing Class II Medical devices

Recall started
12 September 2025
Reason
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
Distribution
US Nationwide distribution in the states of CA, IA, TN, TX, VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0165-2026, Z-0166-2026

Ongoing Class II Medical devices

Recall started
13 August 2025
Reason
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauriti
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2598-2025, Z-2601-2025, Z-2597-2025, Z-2603-2025, Z-2604-2025, Z-2596-2025, Z-2595-2025, Z-2600-2025, Z-2602-2025, Z-2599-2025

Ongoing Class II Medical devices

Recall started
30 June 2025
Reason
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cu
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2211-2025, Z-2215-2025, Z-2205-2025, Z-2206-2025, Z-2213-2025, Z-2208-2025, Z-2200-2025, Z-2218-2025, Z-2207-2025, Z-2210-2025, Z-2201-2025, Z-2217-2025 and 7 more

Ongoing Class II Medical devices

Recall started
29 May 2025
Reason
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1931-2025, Z-1932-2025, Z-1936-2025, Z-1929-2025, Z-1930-2025, Z-1935-2025, Z-1933-2025, Z-1934-2025, Z-1928-2025

Ongoing Class II Medical devices

Recall started
29 May 2025
Reason
Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.
Distribution
Domestic: AZ, CA, CO, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Lithuania, Macao, Malaysia, Netherlands, New Zeal
Initiated
Voluntary: Firm initiated
Recall number
Z-2034-2025

Ongoing Class II Medical devices

Recall started
13 May 2025
Reason
Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
Distribution
Worldwide distribution - US Nationwide and the countries of Andorra, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Denmark, Finland, France, French Guiana, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Iceland, Indonesia, Iran, Ire
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1794-2025, Z-1800-2025, Z-1799-2025, Z-1805-2025, Z-1803-2025, Z-1797-2025, Z-1796-2025, Z-1789-2025, Z-1801-2025, Z-1802-2025, Z-1791-2025, Z-1798-2025 and 6 more

Ongoing Class II Medical devices

Recall started
7 April 2025
Reason
Disconnection of the adapter from the endotracheal tube can potentially be difficult or not possible.
Distribution
Worldwide distribution - US Nationwide and the countries of AE, AL, AO, AR, AT, AU, AW, BD, BE, BG, BH, BO, BR, BW, BY, CA, CD, CH, CI, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, ET, FI, FJ, FO, FR, GA, GB, GF, GH, GI, GP, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KE,
Initiated
Voluntary: Firm initiated
Recall number
Z-1722-2025

Ongoing Class II Medical devices

Recall started
1 April 2025
Reason
Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.
Distribution
Worldwide distribution: US (Nationwide); and OUS (Foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Czech Republic, Estonia, France, Germany, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Lebanon, Libya, Luxembourg,
Initiated
Voluntary: Firm initiated
Recall number
Z-1660-2025

Ongoing Class II Medical devices

Recall started
10 March 2025
Reason
Potential for collimator to fall as a result of incorrect installation.
Distribution
Worldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federatio
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1554-2025, Z-1553-2025

Ongoing Class II Medical devices

Recall started
25 February 2025
Reason
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
Distribution
US
Initiated
FDA Mandated
Recall numbers
Z-1581-2025, Z-1582-2025, Z-1580-2025

Ongoing Class II Medical devices

Recall started
15 January 2025
Reason
Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.
Distribution
US distribution to Florida, North Carolina, and Oregan. International distribution to Canada, Estonia, France, Germany, Japan, Netherlands, Spain, Switzerland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1107-2025, Z-1108-2025, Z-1110-2025, Z-1109-2025, Z-1106-2025

Ongoing Class II Medical devices

Recall started
20 November 2024
Reason
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
Distribution
Nationwide distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-0686-2025

Ongoing Class II Medical devices

Recall started
16 November 2024
Reason
Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, Israel, Italy, Japan, Jordan, Macao, Malaysia, Net
Initiated
Voluntary: Firm initiated
Recall number
Z-0583-2025

Ongoing Class II Medical devices

Recall started
4 November 2024
Reason
Potential safety issue where a patient may be harmed while preparing for or during a scan.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0537-2025, Z-0544-2025, Z-0541-2025, Z-0540-2025, Z-0548-2025, Z-0545-2025, Z-0546-2025, Z-0539-2025, Z-0538-2025, Z-0547-2025, Z-0542-2025, Z-0543-2025 and 1 more

Ongoing Class II Medical devices

Recall started
3 September 2024
Reason
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldiv
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3195-2024, Z-3197-2024, Z-3196-2024, Z-3193-2024, Z-3204-2024, Z-3200-2024, Z-3203-2024, Z-3198-2024, Z-3194-2024, Z-3205-2024, Z-3199-2024, Z-3201-2024 and 1 more

Ongoing Class II Medical devices

Recall started
30 August 2024
Reason
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-3249-2024

Ongoing Class II Medical devices

Recall started
28 June 2024
Reason
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
Distribution
Nationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China Colombia Denmark France Germany Hong Kong India Indonesia Italy Japan Netherlands Portugal Singapore South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand United Kingdom Vietnam
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2510-2024, Z-2508-2024, Z-2509-2024, Z-2511-2024

Ongoing Class I Medical devices

Recall started
5 June 2024
Reason
Potential for coils to heat up and harm patients (burn).
Distribution
Worldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Ko
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2163-2024, Z-2164-2024, Z-2165-2024

Ongoing Class II Medical devices

Recall started
22 May 2024
Reason
Potential for inaccurate fetal heart rate measurements when monitoring multiples.
Distribution
Worldwide distribution - US Nationwide and the countries of AD, AE, AR, AT, AU, BA, BE, BG, BH, BO, BR, CH, CL, CO, CY, CZ, DE, DZ, EG, ES, FI, FR, GB, GI, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JP, KE, KW, LB, LK, LS, LT, LU, LV, LY, MA, ME, MT, MU, MY, NA, NI, NL, NZ, OM, PA, PH, PK, PL, PR,
Initiated
Voluntary: Firm initiated
Recall number
Z-2239-2024

Ongoing Class II Medical devices

Recall started
3 May 2024
Reason
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1839-2024, Z-1857-2024, Z-1850-2024, Z-1851-2024, Z-1861-2024, Z-1844-2024, Z-1843-2024, Z-1854-2024, Z-1859-2024, Z-1864-2024, Z-1853-2024, Z-1838-2024 and 18 more

Ongoing Class II Medical devices

Recall started
3 May 2024
Reason
Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, J
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1950-2024, Z-1949-2024, Z-1948-2024

Ongoing Class II Medical devices

Recall started
22 April 2024
Reason
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
Distribution
Worldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1829-2024, Z-1830-2024

Ongoing Class II Medical devices

Recall started
19 April 2024
Reason
Event Catalog information does not save when copied and transferred from one unit to another.
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, MA, MI, MN, MO, NJ, NM, NY, OH, PA, TN, TX, WA, WV, & DC. The countries of AE, AT, AU, BE, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, IE, IL, IN, IT, JP, KR, LT, NL, NO, OM, PL, PT, RO, SE, SG, SK, TH, & ZA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1827-2024

Ongoing Class II Medical devices

Recall started
15 April 2024
Reason
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1872-2024, Z-1871-2024

Ongoing Class II Medical devices

Recall started
29 March 2024
Reason
Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
Distribution
Worldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1756-2024

Ongoing Class II Medical devices

Recall started
22 March 2024
Reason
Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Kaza
Initiated
Voluntary: Firm initiated
Recall number
Z-1690-2024

Ongoing Class II Medical devices

Recall started
12 March 2024
Reason
Patient support table floor plate may be incorrectly installed.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1565-2024, Z-1550-2024, Z-1581-2024, Z-1553-2024, Z-1562-2024, Z-1575-2024, Z-1577-2024, Z-1576-2024, Z-1573-2024, Z-1559-2024, Z-1556-2024, Z-1578-2024 and 29 more

Ongoing Class II Medical devices

Recall started
28 February 2024
Reason
It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1387-2024, Z-1386-2024

Ongoing Class II Medical devices

Recall started
15 February 2024
Reason
If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a patient condition which prompted the alarm to generate. Although unlikely, this could potentially result in a patient harm.
Distribution
Worldwide - US Nationwide distribution in the states of CA, IL, ME, NC, NJ, NM, NY, OH, VA, WA and the countries of Australia, Austria, Canada, France, Germany, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-3312-2024

Ongoing Class II Medical devices

Recall started
3 January 2024
Reason
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0952-2024, Z-0956-2024, Z-0951-2024, Z-0945-2024, Z-0949-2024, Z-0947-2024, Z-0948-2024, Z-0950-2024, Z-0954-2024, Z-0946-2024, Z-0953-2024, Z-0955-2024

Ongoing Class III Medical devices

Recall started
29 December 2023
Reason
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Distribution
Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0809-2024, Z-0812-2024, Z-0811-2024, Z-0807-2024, Z-0808-2024, Z-0810-2024

Ongoing Class I Medical devices

Recall started
15 December 2023
Reason
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
Distribution
Domestic distribution nationwide. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0929-2024, Z-0931-2024, Z-0930-2024

Ongoing Class II Medical devices

Recall started
15 December 2023
Reason
Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)
Distribution
Nationwide Foreign: Australia Germany India Italy Malaysia Panama Thailand
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1073-2024, Z-1072-2024

Ongoing Class II Medical devices

Recall started
7 December 2023
Reason
If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.
Distribution
Domestic distribution to the following states: AL CA CO FL GA HI IL IN KS KY LA MD MI MO MS ND NE NJ NY OH OK PA RI SC TX UT VT WA International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0719-2024

Ongoing Class II Medical devices

Recall started
30 November 2023
Reason
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0678-2024, Z-0677-2024, Z-0679-2024

Ongoing Class II Medical devices

Recall started
30 November 2023
Reason
Multiple software issues that affect device functionality.
Distribution
Domestic: Nationwide Distribution; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Panama, Palestine, Portugal, South Korea, Spain, Switzerland, Thailand, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0725-2024

Ongoing Class II Medical devices

Recall started
28 November 2023
Reason
Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.
Distribution
Worldwide distribution: US (Nationwide) to states of: CA, GA, IA, MA, MI, MN, NJ, NM, NY, OH, PA, TN, TX, WA, WV; and OUS (Foreign) distribution to countries of: Australia, Austria, Belarus, Belgium, Brazil, Canada, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, India,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0731-2024, Z-0732-2024

Ongoing Class I Medical devices

Recall started
17 November 2023
Reason
Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.
Distribution
Domestic distribution to AL AZ CA CO DE FL GA IA IL IN KS KY LA MA MD MI MO NE NJ NV NY OH OK OR PA SC TN TX UT VA VT WA WI WV International distribution to Argentina Austria Canada China Colombia Croatia Czech Republic Denmark Ecuador Egypt France Germany Greece Italy Kuwait Libya Netherlands NO93
Initiated
Voluntary: Firm initiated
Recall number
Z-0505-2024

Ongoing Class II Medical devices

Recall started
23 October 2023
Reason
Potential component failure in the Gradient Coil could product smoke and/or fire.
Distribution
Domestic: Nationwide Distribution; Foreign: Albania, Andorra, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, I
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0395-2024, Z-0392-2024, Z-0389-2024, Z-0391-2024, Z-0390-2024, Z-0394-2024, Z-0393-2024

Ongoing Class II Medical devices

Recall started
9 October 2023
Reason
This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).
Distribution
US Nationwide distribution in the state of MN.
Initiated
Voluntary: Firm initiated
Recall number
Z-0319-2024

Ongoing Class II Medical devices

Recall started
9 October 2023
Reason
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
Distribution
Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0293-2024, Z-0292-2024, Z-0294-2024

Ongoing Class II Medical devices

Recall started
9 October 2023
Reason
When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.
Distribution
US Nationwide distribution in the state of MN.
Initiated
Voluntary: Firm initiated
Recall number
Z-0306-2024

Ongoing Class II Medical devices

Recall started
4 October 2023
Reason
Philips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
Distribution
United States Territories of Puerto Rico, Guam, and Virgin Islands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0228-2024, Z-0224-2024, Z-0230-2024, Z-0226-2024, Z-0229-2024, Z-0227-2024

Ongoing Class II Medical devices

Recall started
21 September 2023
Reason
Potential for units suspended on the ceiling to fail and fall.
Distribution
Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0272-2024, Z-0271-2024, Z-0274-2024, Z-0277-2024, Z-0278-2024, Z-0280-2024, Z-0276-2024, Z-0275-2024, Z-0279-2024, Z-0273-2024

Ongoing Class II Medical devices

Recall started
31 August 2023
Reason
A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Distribution
Worldwide - US Nationwide distribution in the states of MA, MI. The country of Brazil.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2654-2023, Z-2655-2023

Ongoing Class II Medical devices

Recall started
17 August 2023
Reason
Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients
Distribution
Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland Portugal Qatar Romania Slovakia South Africa South Korea Spain Sweden Sweden Switzerland United Kingdom Ut
Initiated
Voluntary: Firm initiated
Recall number
Z-2608-2023

Ongoing Class II Medical devices

Recall started
14 June 2023
Reason
Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Without Option C01 Full Arrhythmia, the device will not provide the yellow alarms for enhanced arrhythmia detection. This is a retrospectively reported recall from 6/14/23.
Distribution
Foreign Only: Denmark France Germany Japan United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-0020-2025

Ongoing Class II Medical devices

Recall started
12 June 2023
Reason
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2398-2023, Z-2400-2023, Z-2399-2023

Ongoing Class II Medical devices

Recall started
12 June 2023
Reason
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2395-2023, Z-2397-2023, Z-2396-2023

Ongoing Class II Medical devices

Recall started
19 May 2023
Reason
There is a software login in issue that may prevent the user from logging in.
Distribution
Distribution in US: AZ, GA, IN, IA, MN, NE, NH, OH, TX OUS distribution to Australia, Chile, China, Malta, Namibia, New Zealand, Oman, Panama, Qatar, Suriname, United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-2042-2023

Ongoing Class II Medical devices

Recall started
25 April 2023
Reason
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
Distribution
Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Mace
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2259-2023, Z-2258-2023, Z-2260-2023, Z-2261-2023

Ongoing Class II Medical devices

Recall started
19 April 2023
Reason
During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0035-2025

Ongoing Class II Medical devices

Recall started
20 March 2023
Reason
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, MI, MO, NY, TN, TX, VT and the countries of Australia, Chile, Costa Rica, Czech Republic, Dominican, Republic Germany, India, Israel, Japan, Lithuania, Philippines, Poland, Sint Maarten (Dutch part), South Africa, Thailand,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1586-2023, Z-1585-2023

Ongoing Class II Medical devices

Recall started
17 March 2023
Reason
Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operators foot. If foot or leg entrapment occurs while pressing Unload Pedal, the risk to operators may include: pain, muscle or ligament sprain, abrasion, laceration, contusion, fracture, loss of function, and joint dislocation.
Distribution
Worldwide - US Nationwide distribution including in the states of States: AZ, IN, MA, MD, MN, NY, OH, PA, and TX. The countries of Australia, Austria, China, Czech Republic, Denmark, France, Germany, Israel, Japan, Malaysia, Netherlands, Norway, Palestine, Panama, South Korea, Spain, Switzerland, Th
Initiated
Voluntary: Firm initiated
Recall number
Z-1584-2023

Ongoing Class II Medical devices

Recall started
9 March 2023
Reason
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users
Distribution
Nationwide including Puerto Rico and Virgin Islands Foreign: ARGENTINA Australia Austria Bosnia and Herzegovina Bulgaria China Colombia Cyprus Czech Republic Egypt France French Polynesia Germany Greece Hungary Indonesia Italy Jordan Kenya Latvia Lithuania Macedonia Malta Mexico Netherlands New Zeal
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1382-2023, Z-1383-2023

Ongoing Class II Medical devices

Recall started
6 March 2023
Reason
EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.
Distribution
Nationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1644-2023, Z-1645-2023

Ongoing Class II Medical devices

Recall started
3 March 2023
Reason
Philips has identified a compliance issue where ProxiDiagnost N90 R.1.0, CombiDiagnost R90 R.1.0 and CombiDiagnost R90 R1.1 systems are missing a certification label required by the Food and Drug Administration (FDA).
Distribution
US Nationwide
Initiated
FDA Mandated
Recall numbers
Z-1587-2023, Z-1588-2023, Z-1589-2023

Ongoing Class II Medical devices

Recall started
18 January 2023
Reason
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0007-2025

Ongoing Class II Medical devices

Recall started
22 November 2022
Reason
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0461-2023, Z-0462-2023, Z-0459-2023, Z-0460-2023, Z-0463-2023

Ongoing Class II Medical devices

Recall started
21 November 2022
Reason
Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of this product.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Croatia, Cura¿ao, Czechia, Denmark, Faroe Islands, Finland, France, Germany, Hong Kong, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Luxembourg, Malaysi
Initiated
Voluntary: Firm initiated
Recall number
Z-0908-2023

Ongoing Class II Medical devices

Recall started
21 October 2022
Reason
Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Chile, China, Germany, India, Italy, Malaysia, Netherlands, Panama, Spain, Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1477-2023

Ongoing Class II Medical devices

Recall started
6 September 2022
Reason
Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.
Distribution
Domestic distribution to AL AZ CA CO FL GA ID IL IN LA MA MI MN NC NE NJ OH PA SC TN TX UT VT WA WY. Foreign distribution to Australia Austria Belgium Czech Republic Denmark Estonia Finland France Germany Ireland Italy Malta Netherlands Norway Poland Portugal Qatar Singapore Slovakia Spain Sweden S
Initiated
Voluntary: Firm initiated
Recall number
Z-0004-2023

Completed Class II Medical devices

Recall started
29 March 2022
Reason
Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use.
Distribution
Domestic: AL, AR, AZ, CA, FL, GA, HI, IL, IN, KY, LA, MD, ME, MI, MS, NC, NJ, NM, NY, OH, OK, PA, RI, TN, TX; Foreign: Albania, Argentina, Australia, Austria, Bolivia, Bosnia-Herzegovina , Cambodia, Canada, Chile, Colombia, Costa Rica, Czech Republic, Ecuador, Egypt, Ethiopia, France, Germany, Greec
Initiated
Voluntary: Firm initiated
Recall number
Z-0576-2024

Ongoing Class II Medical devices

Recall started
25 February 2022
Reason
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Added 1/3/2023), Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
Distribution
Nationwide Distribution including Puerto Rico
Initiated
FDA Mandated
Recall number
Z-1563-2022

Terminated Class II Medical devices

Recall started
22 February 2022
Terminated
6 November 2024
Reason
When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.
Distribution
Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1041-2022

Ongoing Class II Medical devices

Recall started
7 February 2022
Reason
Update to instructions for use regarding the cleaning and disinfection process for the reusable probes.
Distribution
Worldwide distribution: US (nationwide) Including Puerto Rico and OUS (foreign) countries including: Afghanistan, Albania, Algeria, Angola, Argentina, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Da
Initiated
Voluntary: Firm initiated
Recall number
Z-0926-2022

Ongoing Class II Medical devices

Recall started
4 February 2022
Reason
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.
Distribution
Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Malaysia, Switzerland, Chile, Ireland, Norway, Hong Kong, Argentina, Mexico, Bolivia,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0881-2022, Z-0882-2022

Ongoing Class II Medical devices

Recall started
2 February 2022
Reason
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Distribution
Global Distribution. US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0874-2022

Ongoing Class II Medical devices

Recall started
6 January 2022
Reason
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Distribution
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0975-2024, Z-0973-2024, Z-0974-2024

Terminated Class II Medical devices

Recall started
3 January 2022
Terminated
11 April 2024
Reason
eCareManager (eCM) Sentry Score software not approved for use
Distribution
Worldwide distribution - US Nationwide and the country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0549-2022

Ongoing Class II Medical devices

Recall started
29 November 2021
Reason
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
Distribution
US Nationwide and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0509-2022, Z-0507-2022, Z-0506-2022, Z-0508-2022

Ongoing Class II Medical devices

Recall started
26 November 2021
Reason
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
Distribution
US Nationwide and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-0505-2022

Ongoing Class II Medical devices

Recall started
24 November 2021
Reason
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Distribution
US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sr
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0724-2022, Z-0725-2022

Terminated Class II Medical devices

Recall started
19 November 2021
Terminated
7 February 2024
Reason
Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.
Distribution
Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0416-2022

Ongoing Class II Medical devices

Recall started
12 November 2021
Reason
Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Algeria, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cameroon, Chile, China, Colombia, Cook Islands, Costa Rica, Croatia, Cuba, Cyprus, Czech Repu
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0355-2022, Z-0357-2022, Z-0356-2022

Terminated Class II Medical devices

Recall started
12 November 2021
Terminated
5 February 2024
Reason
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available
Distribution
Worldwide distribution - US Nationwide distribution in the state of WA and the countries of France, Germany, Japan, Poland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0471-2022

Ongoing Class II Medical devices

Recall started
29 October 2021
Reason
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Bangladesh, Belgium, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Ital
Initiated
Voluntary: Firm initiated
Recall number
Z-0476-2022

Ongoing Class II Medical devices

Recall started
21 October 2021
Reason
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1355-2023, Z-1356-2023

Terminated Class II Medical devices

Recall started
24 September 2021
Terminated
26 September 2024
Reason
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Distribution
Domestic distribution nationwide. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0213-2022

Terminated Class II Medical devices

Recall started
13 August 2021
Terminated
11 March 2024
Reason
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Distribution
US Nationwide distribution in the states of ID, MA, MI, TX, VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2474-2021, Z-2476-2021, Z-2475-2021

Completed Class II Medical devices

Recall started
11 August 2021
Reason
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2414-2021, Z-2421-2021, Z-2422-2021, Z-2411-2021, Z-2420-2021, Z-2419-2021, Z-2416-2021, Z-2412-2021, Z-2417-2021, Z-2418-2021, Z-2413-2021, Z-2415-2021

Terminated Class II Medical devices

Recall started
30 July 2021
Terminated
25 July 2023
Reason
Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Neth
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2344-2021, Z-2342-2021, Z-2346-2021, Z-2341-2021, Z-2343-2021, Z-2345-2021, Z-2347-2021

Terminated Class II Medical devices

Recall started
16 July 2021
Terminated
12 May 2023
Reason
Infant/Child defibrillator pads contain incorrect labeling. This could lead to a delay in therapy, or delivery of lower energy than indicated for an adult patient
Distribution
Worldwide distribution. US Nationwide, Australia, Austria, Bolivia, Canada, Chile, China, Denmark, France, French Polynesia, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Republic of Korea, Malta, Martinique, Mexico, Monaco, Netherlands, Norway, Poland, Spain, Switzerland, Taiwan, United Kingdo
Initiated
Voluntary: Firm initiated
Recall number
Z-2427-2021

Ongoing Class II Medical devices

Recall started
15 July 2021
Reason
The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
Distribution
Domestic distribution to the following states and territories: CA DE FL GA IL KY MA ME MI MO NH NJ NM NY OH PA SC TN TX VA VT WI WY, and PR. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0715-2024

Terminated Class II Medical devices

Recall started
8 June 2021
Terminated
10 August 2023
Reason
When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
Distribution
US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2318-2021, Z-2317-2021, Z-2316-2021, Z-2314-2021, Z-2315-2021

Ongoing Class II Medical devices

Recall started
1 May 2021
Reason
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1985-2021, Z-1988-2021, Z-1979-2021, Z-1986-2021, Z-1982-2021, Z-1981-2021, Z-1980-2021, Z-1984-2021, Z-1987-2021, Z-1983-2021, Z-1990-2021, Z-1989-2021

Ongoing Class II Medical devices

Recall started
1 May 2021
Reason
Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part
Distribution
US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of: Australia, Germany, Netherlands and Saudi Arabia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0281-2022, Z-0278-2022, Z-0280-2022, Z-0279-2022, Z-0277-2022

Ongoing Class II Medical devices

Recall started
23 April 2021
Reason
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
Distribution
Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1795-2021, Z-1796-2021

Terminated Class II Medical devices

Recall started
13 April 2021
Terminated
23 September 2022
Reason
Cooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injury
Distribution
Worldwide distribution - US Nationwide distribution in the state of NV and the country of Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1774-2021

Ongoing Class II Medical devices

Recall started
29 March 2021
Reason
Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor
Distribution
Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Bulgaria Canada Chile China Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Finland France Germany Greece Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy J
Initiated
Voluntary: Firm initiated
Recall number
Z-1517-2021

Terminated Class II Medical devices

Recall started
23 March 2021
Terminated
28 September 2023
Reason
Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit
Distribution
International distribution in the country of China.
Initiated
Voluntary: Firm initiated
Recall number
Z-1326-2021

Terminated Class II Medical devices

Recall started
22 February 2021
Terminated
9 February 2023
Reason
Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees
Distribution
US Virgin Island Foreign: Australia Bolivia China Germany Hungary Indonesia Italy Japan Korea Korea, Republic of Latvia Palestine Philippines Romania Russian Federation Senegal Thailand Uzbekistan
Initiated
Voluntary: Firm initiated
Recall number
Z-1265-2021

Ongoing Class II Medical devices

Recall started
9 February 2021
Reason
Customers were not notified of previous recalls associated with various defibrillator models.
Distribution
U.S.: CA, CT, FL, GA, IN, MI, MO, NJ, NY, OH, TX, and WA. International distribution in the countries of Australia, Canada, France, Germany, Japan, Norway, Switzerland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0081-2022, Z-0079-2022, Z-0078-2022, Z-0080-2022

Terminated Class II Medical devices

Recall started
2 February 2021
Terminated
20 September 2022
Reason
Manufacturing failed to follow process steps resulting in the distribution of potentially defective devices
Distribution
Worldwide distribution - US Nationwide distribution in the states of IN, MN and the country of Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-1671-2021

Ongoing Class II Medical devices

Recall started
24 December 2020
Reason
Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured
Distribution
Nationwide including Puerto Rico Foreign: Albania Argentina Australia Austria Bangladesh Bolivia Brazil Bulgaria China Czech Republic Denmark Ecuador Egypt France Germany Greece Hungary India Indonesia Iraq Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kuwait Latvia Lebanon Libya Malaysia M
Initiated
Voluntary: Firm initiated
Recall number
Z-1507-2024

Terminated Class II Medical devices

Recall started
19 November 2020
Terminated
19 April 2023
Reason
A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Isra
Initiated
Voluntary: Firm initiated
Recall number
Z-0852-2021

Terminated Class II Medical devices

Recall started
18 November 2020
Terminated
22 July 2021
Reason
Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.
Distribution
Worldwide distribution - US Nationwide distribution in the state of SD and the countries of China, England, Germany, Hong Kong, Portugal.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0545-2021, Z-0544-2021

Terminated Class II Medical devices

Recall started
14 September 2020
Terminated
13 September 2022
Reason
The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
Distribution
Global Distribution. US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0284-2021, Z-0285-2021

Terminated Class II Medical devices

Recall started
16 July 2020
Terminated
3 August 2021
Reason
The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user. These device behaviors could result in a delay in therapy or a failure to deliver the intended therapy.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2794-2020

Terminated Class II Medical devices

Recall started
30 June 2020
Terminated
25 January 2024
Reason
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Distribution
Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2696-2020, Z-2697-2020

Ongoing Class II Medical devices

Recall started
25 June 2020
Reason
The device may experience an interruption of gas measurement due to a firmware issue, ceasing measurement and display of gas levels and generating one or more technical or INOP alarms.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2689-2020

Terminated Class II Medical devices

Recall started
18 June 2020
Terminated
10 January 2023
Reason
A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
Distribution
World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2812-2020, Z-2810-2020, Z-2811-2020

Terminated Class II Medical devices

Recall started
8 June 2020
Terminated
9 January 2023
Reason
Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor.
Distribution
Nationwide Foreign: Angola Angola Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana Germany Ghana Gibraltar Greece Greenlan
Initiated
Voluntary: Firm initiated
Recall number
Z-2531-2020

Terminated Class II Medical devices

Recall started
2 June 2020
Terminated
9 February 2021
Reason
Thermo switches in the main power supply for the system may be incorrectly installed, resulting in a malfunction.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CT, FL, GA, HI, IL, LA, MA, MD, MI, MS, MT, NC, NY, TN, TX and the country of Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-2523-2020

Terminated Class II Medical devices

Recall started
19 May 2020
Terminated
25 August 2020
Reason
Philips has identified a Philips SureSigns VM1 monitor in the possession of a clinical user that was not intended for commercial distribution and for which Philips does not have records regarding its manufacture. Philips is therefore removing this device from distribution.
Distribution
US Nationwide distribution including in the state of NJ.
Initiated
Voluntary: Firm initiated
Recall number
Z-2337-2020

Terminated Class II Medical devices

Recall started
27 April 2020
Terminated
19 July 2021
Reason
HeartStart XL MonitorlDefibrillators may not have been included in prior field actions and received required corrections
Distribution
Worldwide distribution - US Nationwide distribution and the countries of China, Republic of Korea, Netherlands, Norway, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2157-2020

Terminated Class II Medical devices

Recall started
23 April 2020
Terminated
30 March 2023
Reason
The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.
Distribution
Domestic distribution nationwide. Foreign distribution to Chile and Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-2256-2020

Terminated Class II Medical devices

Recall started
21 April 2020
Terminated
16 July 2021
Reason
HeartStart XL MonitorlDefibrillators may not have been included in prior field actions and received required corrections
Distribution
Worldwide distribution - US Nationwide distribution including in the states of KY, MA, ND, NY and the countries of United Kingdom, Bahrain.
Initiated
Voluntary: Firm initiated
Recall number
Z-2137-2020

Terminated Class II Medical devices

Recall started
20 April 2020
Terminated
2 April 2021
Reason
Internal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instructions for Use (IFU) to perform routine operational checks are followed. The instructions for use rely on these checks rather than specifying a maximum number of sterilization cycles to determine when the paddles should be discarded. A damaged Internal Paddle may not be able to deliver therapy.
Distribution
Worldwide distribution - US Nationwide distrubtion and the countries of Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Congo (Democratic Republic of the), Costa Rica, C¿te D'Ivoire, Cyprus, Czec
Initiated
Voluntary: Firm initiated
Recall number
Z-2134-2020

Terminated Class II Medical devices

Recall started
13 April 2020
Terminated
7 April 2021
Reason
The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2053-2020

Terminated Class II Medical devices

Recall started
13 April 2020
Terminated
14 April 2023
Reason
The labeling of the Philips Chest Pneumograph does not include a statement indicating that the product contains natural rubber latex.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2245-2020, Z-2244-2020

Terminated Class II Medical devices

Recall started
10 April 2020
Terminated
11 January 2021
Reason
When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1933-2020, Z-1932-2020

Terminated Class II Medical devices

Recall started
10 April 2020
Terminated
10 February 2021
Reason
Damaged ESD bags used for storing components in Processor PCA Replacement Kits
Distribution
Worldwide distribution - US Nationwide distribution including in the states of OK, NY and the countries of Libya, Netherlands, Canada, South Africa, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2064-2020

Terminated Class II Medical devices

Recall started
7 April 2020
Terminated
28 April 2023
Reason
If the shaft breaks due to material fatigue, the trauma arm can fall down and may result in injury.
Distribution
Worldwide distribution - US Nationwide distribution including the states of CA, ME, RI, TX, WI and the countries of Canada, Australia, Austria, France, Germany, Greece, Hungary, Ireland, Netherlands, Portugal, Slovenia, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1806-2020

Terminated Class II Medical devices

Recall started
6 April 2020
Terminated
16 September 2022
Reason
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
Distribution
Worldwide distribution including one unit distributed to Iowa.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1922-2020, Z-1921-2020, Z-1923-2020

Terminated Class II Medical devices

Recall started
3 April 2020
Terminated
1 February 2023
Reason
When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.
Distribution
Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand
Initiated
Voluntary: Firm initiated
Recall number
Z-1809-2020

Terminated Class II Medical devices

Recall started
5 March 2020
Terminated
9 September 2022
Reason
No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..
Distribution
Worldwide distribution: US (nationwide) including the following states: Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio,
Initiated
Voluntary: Firm initiated
Recall number
Z-1801-2020

Terminated Class II Medical devices

Recall started
5 March 2020
Terminated
11 February 2021
Reason
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
Distribution
The products were distributed US nationwide. Global distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1767-2020, Z-1766-2020

Terminated Class II Medical devices

Recall started
26 February 2020
Terminated
8 December 2022
Reason
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports involved a death of a patient who could not be resuscitated.
Distribution
US Nationwide domestic distribution. Worldwide foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1499-2020

Terminated Class II Medical devices

Recall started
20 February 2020
Terminated
21 September 2021
Reason
Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00
Distribution
Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary,
Initiated
Voluntary: Firm initiated
Recall number
Z-1616-2020

Terminated Class II Medical devices

Recall started
14 January 2020
Terminated
9 February 2021
Reason
M4758A Ambulance Parameter Extension (APE) kit may cause ECG "Leads Off" condition in hot, humid environments
Distribution
International distribution in the countries of Austria, Germany, Netherlands, Norway, Poland, Switzerland
Initiated
Voluntary: Firm initiated
Recall number
Z-1342-2020

Terminated Class II Medical devices

Recall started
13 January 2020
Terminated
22 June 2023
Reason
if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.
Distribution
US Nationwide Foreign: Afghanistan Albania Algeria Argentina Armenia Aruba Australia Austria Azerbaijan Bangladesh Belarus Belgium Bermuda Bulgaria Burkina Faso Canada China C¿te D'Ivoire Croatia Curacao Cyprus Czech Republic Egypt France Germany Ghana Gibraltar Greece
Initiated
Voluntary: Firm initiated
Recall number
Z-1345-2020

Terminated Class II Medical devices

Recall started
30 December 2019
Terminated
17 November 2022
Reason
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1488-2020, Z-1494-2020, Z-1492-2020, Z-1490-2020, Z-1487-2020, Z-1489-2020, Z-1491-2020, Z-1493-2020, Z-1486-2020

Terminated Class II Medical devices

Recall started
23 December 2019
Terminated
17 November 2022
Reason
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
Distribution
Worldwide Distribution. US nationwide, Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Latvia, Libya, Mexico, Netherland
Initiated
Voluntary: Firm initiated
Recall number
Z-1145-2020

Terminated Class II Medical devices

Recall started
20 November 2019
Terminated
4 April 2024
Reason
Unexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.
Distribution
United States
Initiated
FDA Mandated
Recall number
Z-3023-2020

Terminated Class II Medical devices

Recall started
16 October 2019
Terminated
29 November 2023
Reason
The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock with an energy level different from the setting selected by the user.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0498-2020

Terminated Class II Medical devices

Recall started
3 October 2019
Terminated
19 April 2023
Reason
Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0608-2020

Terminated Class II Medical devices

Recall started
10 September 2019
Terminated
29 April 2024
Reason
Affected devices contain a Printed Circuit Assembly (PCA) that was not manufactured/repaired in accordance to quality standards.
Distribution
HeartStart FRx: US - CA, CT, FL, IL, MO, NY,TN, TX and WI OUS - Australia, Brazil, Canada, Germany, Italy, Japan, South Korea, Norway, Switzerland and Taiwan. HeartStart HS1: US - CT, FL, IL, MI, MS, NJ, TN, TX, WA and WI OUS - Argentina, Australia, Canada, France, Germany, Hong Kong, Italy,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1272-2020, Z-1271-2020

Terminated Class II Medical devices

Recall started
23 August 2019
Terminated
3 January 2024
Reason
Remote Antenna Cable Used with Philips MX4O Monitors Does Not Have the Required Rating for Some Installations
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2547-2019

Terminated Class II Medical devices

Recall started
19 July 2019
Terminated
3 June 2020
Reason
The thickness of the connector to the trunk cable on affected limb lead sets may be greater than specified. which may prevent or impede the performance of I 2-Lead ECG measurements.
Distribution
Worldwide Distribution - US Nationwide State KY AZ CA UT OH NY SC CO FL IN TX ND KS GA Canada Germany
Initiated
Voluntary: Firm initiated
Recall number
Z-2495-2019

Terminated Class II Medical devices

Recall started
16 July 2019
Terminated
4 June 2020
Reason
The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.
Distribution
US Nationwide Distribution Foreign: Canada Argentina AUSTRALIA Austria Azerbaijan Bahrain Bangladesh Belgium Bosnia and Herzegovina Brazil Bulgaria Chile Colombia CROATIA Denmark ECUADOR Egypt Finland France Germany Greece Hong Kong India Indonesia Ireland Israel Italie
Initiated
Voluntary: Firm initiated
Recall number
Z-2205-2019

Terminated Class II Medical devices

Recall started
16 July 2019
Terminated
20 October 2022
Reason
Non-conforming devices are identified, which may not have been included in prior field actions and may not have had a number of corrections applicable to these devices.
Distribution
US Nationwide distribution in the state of MA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2488-2019

Terminated Class II Medical devices

Recall started
7 June 2019
Terminated
8 May 2020
Reason
The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery.
Distribution
Nationwide domestic distribution. Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1923-2019

Terminated Class III Medical devices

Recall started
29 April 2019
Terminated
22 November 2024
Reason
Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values
Distribution
Worldwide Distribution - US Nationwide Foreign: Andorra Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Czech Republic Denmark Dominican Rep Egypt Estonia Finland France French Gui
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1772-2019, Z-1771-2019

Ongoing Class II Medical devices

Recall started
19 April 2019
Reason
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1318-2019, Z-1321-2019, Z-1322-2019, Z-1319-2019, Z-1324-2019, Z-1320-2019, Z-1317-2019, Z-1323-2019

Ongoing Class II Medical devices

Recall started
28 March 2019
Reason
The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .
Distribution
Worldwide distribution - US Nationwide and to 71 countries.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1693-2019, Z-1692-2019, Z-1694-2019, Z-1695-2019

Terminated Class II Medical devices

Recall started
12 March 2019
Terminated
7 October 2020
Reason
During manufacturing of an Azurion R2.0 system, it was found out that the cable set was not according to specification. This resulted in damage of the 24V power cable of the detector.
Distribution
Distributed to one account in Florida, and one international account in Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-1130-2019

Terminated Class II Medical devices

Recall started
28 February 2019
Terminated
3 August 2020
Reason
The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
Distribution
US Nationwide domestic distribution, and worldwide international distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1052-2019, Z-1054-2019, Z-1055-2019, Z-1056-2019, Z-1053-2019

Terminated Class II Medical devices

Recall started
25 January 2019
Terminated
8 December 2020
Reason
During use of the Philips FSE, it is possible for the metal electrode tip to break off and remain in the scalp of the newborn following labor and delivery.
Distribution
Distributed nationwide within US. Worldwide international distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1250-2019

Terminated Class II Medical devices

Recall started
25 January 2019
Terminated
21 April 2020
Reason
Lithium-ion Rechargeable Batteries for the Philips MX4O Wearable Monitor may have a manufacturing defect that may affect the operation of the monitor
Distribution
US Nationwide distribution in the states of AK, CA, CO, IN,MI, NC, ND, NM, OR, TN, TX and UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-1067-2019

Ongoing Class II Medical devices

Recall started
9 January 2019
Reason
Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0853-2019, Z-0854-2019, Z-0856-2019, Z-0855-2019, Z-0861-2019, Z-0859-2019, Z-0857-2019, Z-0858-2019, Z-0860-2019

Terminated Class II Medical devices

Recall started
23 November 2018
Terminated
29 September 2020
Reason
Affected lithium-ion batteries may contain a defective component (Thermal Cut-off or TCO). Should the component fail, the battery will no longer charge or deliver power to the Monitor/Defibrillator, which may prevent operation of the device if it is not connected to AC or DC power. Should a second, unaffected battery be present in the Monitor/Defibrillator, the battery failure may go unnoticed.
Distribution
U.S.: AZ, WA, PA, UT, NY, CO, MD, AR, TX, MO, MI, FL, VA, MA, GA, TN, NV, OH, CA, NH, KY, KS, OR, WI, NC, IL, LA, NJ, WV, AL, WY, MS Foreign (OUS): United Kingdom, Korea Rep, Singapore, Australia, Taiwan, Sri Lanka, Brazil, France, Switzerland, Germany, Austria, Spain, Italy, Trinidad and Tobago,
Initiated
Voluntary: Firm initiated
Recall number
Z-0603-2019

Terminated Class II Medical devices

Recall started
8 August 2018
Terminated
29 April 2021
Reason
The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child pad cartridge should not be used on an adult because it delivers lower energy (50 J vs 150 J) than the Adult pads cartridge and may therefore not be effective on an adult.
Distribution
US: AK, AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, WY, Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Denmar
Initiated
Voluntary: Firm initiated
Recall number
Z-2549-2019

Terminated Class I Medical devices

Recall started
12 July 2017
Terminated
27 May 2020
Reason
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Distribution
No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-0489-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.