PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls

59 recalls recorded since 2018, most recently on 16 July 2026. 59 are listed as ongoing. Status and classification are as published by the FDA.

Recalls59
Ongoing (FDA)59
Terminated or completed0
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
16 July 2026
Reason
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2932-2026

Ongoing Class II Medical devices

Recall started
16 July 2026
Reason
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2931-2026

Ongoing Class II Medical devices

Recall started
29 June 2026
Reason
Firmware issue potentially leading to loss of geometry movements.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Germany, Greece,
Initiated
Voluntary: Firm initiated
Recall number
Z-2928-2026

Ongoing Class II Medical devices

Recall started
17 June 2026
Reason
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2893-2026, Z-2892-2026, Z-2891-2026

Ongoing Class II Medical devices

Recall started
18 May 2026
Reason
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
Distribution
US and ROW.
Initiated
FDA Mandated
Recall numbers
Z-2411-2026, Z-2412-2026

Ongoing Class II Medical devices

Recall started
30 April 2026
Reason
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czec
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2258-2026, Z-2259-2026

Ongoing Class II Medical devices

Recall started
28 April 2026
Reason
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
Distribution
Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, E
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2227-2026, Z-2229-2026, Z-2228-2026

Ongoing Class II Medical devices

Recall started
28 April 2026
Reason
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, C¿te D'Ivoire, El Salvador, Estonia, Greece, Honduras, Hong Kong, Iceland, Indonesia, Ira
Initiated
Voluntary: Firm initiated
Recall number
Z-2225-2026

Ongoing Class II Medical devices

Recall started
31 March 2026
Reason
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2170-2026

Ongoing Class II Medical devices

Recall started
19 March 2026
Reason
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bnis insted Us Instas instes, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Estonia, Germany, Greece, Indonesia, Iraq, Italy, Japan, Kuwait, Latvia, Ma
Initiated
Voluntary: Firm initiated
Recall number
Z-1805-2026

Ongoing Class II Medical devices

Recall started
5 March 2026
Reason
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Andorra, Angola, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany,
Initiated
Voluntary: Firm initiated
Recall number
Z-1692-2026

Ongoing Class II Medical devices

Recall started
3 March 2026
Reason
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Distribution
Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colo
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1664-2026, Z-1672-2026, Z-1673-2026, Z-1676-2026, Z-1679-2026, Z-1666-2026, Z-1660-2026, Z-1668-2026, Z-1665-2026, Z-1674-2026, Z-1677-2026, Z-1680-2026 and 10 more

Ongoing Class II Medical devices

Recall started
26 January 2026
Reason
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1348-2026, Z-1349-2026, Z-1346-2026, Z-1350-2026, Z-1345-2026, Z-1344-2026, Z-1347-2026, Z-1351-2026

Ongoing Class II Medical devices

Recall started
15 December 2025
Reason
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Ken
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1063-2026, Z-1074-2026, Z-1075-2026, Z-1062-2026, Z-1064-2026, Z-1069-2026, Z-1077-2026, Z-1065-2026, Z-1070-2026, Z-1076-2026, Z-1068-2026, Z-1067-2026 and 4 more

Ongoing Class II Medical devices

Recall started
12 December 2025
Reason
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
Distribution
International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;
Initiated
Voluntary: Firm initiated
Recall number
Z-1113-2026

Ongoing Class II Medical devices

Recall started
12 December 2025
Reason
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chil
Initiated
Voluntary: Firm initiated
Recall number
Z-1091-2026

Ongoing Class II Medical devices

Recall started
8 December 2025
Reason
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia
Initiated
Voluntary: Firm initiated
Recall number
Z-1079-2026

Ongoing Class II Medical devices

Recall started
2 October 2025
Reason
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Distribution
Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines,
Initiated
Voluntary: Firm initiated
Recall number
Z-0719-2026

Ongoing Class II Medical devices

Recall started
29 September 2025
Reason
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
Distribution
Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0241-2026

Ongoing Class II Medical devices

Recall started
3 September 2025
Reason
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Fi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2627-2025, Z-2633-2025, Z-2624-2025, Z-2634-2025, Z-2625-2025, Z-2635-2025, Z-2623-2025, Z-2628-2025, Z-2631-2025, Z-2629-2025, Z-2626-2025, Z-2630-2025 and 1 more

Ongoing Class II Medical devices

Recall started
3 September 2025
Reason
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmar
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0002-2026, Z-0010-2026, Z-0009-2026, Z-0013-2026, Z-0004-2026, Z-0001-2026, Z-0007-2026, Z-0014-2026, Z-0006-2026, Z-0012-2026, Z-0015-2026, Z-0011-2026 and 3 more

Ongoing Class II Medical devices

Recall started
8 August 2025
Reason
Software issue that results in the display of outdated information.
Distribution
US Nationwide distribution in the states of GA, NC & TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2543-2025

Ongoing Class II Medical devices

Recall started
1 August 2025
Reason
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hun
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2502-2025, Z-2503-2025

Ongoing Class II Medical devices

Recall started
1 August 2025
Reason
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
Distribution
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, DC, FL, GA, HI, ID, IL, KS, KY, LA, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI, WV and the countries of Argentina, Austria, Australia, Belgium, Switzerland, Chile, China, Colombia, Czechia, Germany, Egypt, Spain,
Initiated
Voluntary: Firm initiated
Recall number
Z-2504-2025

Ongoing Class II Medical devices

Recall started
30 June 2025
Reason
Potential for temporary loss of imaging (X-ray) functionality due to software issue.
Distribution
Worldwide - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polyn
Initiated
Voluntary: Firm initiated
Recall number
Z-2189-2025

Ongoing Class II Medical devices

Recall started
4 June 2025
Reason
Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
Distribution
Worldwide distribution: US (nationwide) to states of: CA, CO, DC, DE, FL, IL, IN, KY, MA, MD, MN, MO, MT, NJ, NV, NY, OH, SC, TX, WA; and OUS (Foreign) to countries of: Argentina (AR), Australia (AU), Canada (CA), Chile (CL), Germany (DE), Dominican Republic (DO), Egypt (EG), ES, France (FR), United
Initiated
Voluntary: Firm initiated
Recall number
Z-2134-2025

Ongoing Class II Medical devices

Recall started
28 May 2025
Reason
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica
Initiated
Voluntary: Firm initiated
Recall number
Z-1969-2025

Ongoing Class II Medical devices

Recall started
7 April 2025
Reason
An issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.
Distribution
Domestic: Nationwide Distribution International: Australia, Austria, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Indonesia, Ireland, Italy, Japan, Luxembourg, Mexico, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, U
Initiated
Voluntary: Firm initiated
Recall number
Z-1740-2025

Ongoing Class II Medical devices

Recall started
24 March 2025
Reason
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Distribution
US: Nationwide Distribution International: Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican R
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1776-2025, Z-1783-2025, Z-1775-2025, Z-1782-2025, Z-1772-2025, Z-1770-2025, Z-1777-2025, Z-1771-2025, Z-1780-2025, Z-1778-2025, Z-1779-2025, Z-1774-2025 and 2 more

Ongoing Class II Medical devices

Recall started
28 February 2025
Reason
Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.
Distribution
Domestic: MN
Initiated
Voluntary: Firm initiated
Recall number
Z-1324-2025

Ongoing Class II Medical devices

Recall started
13 January 2025
Reason
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, Chin
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1087-2025, Z-1091-2025, Z-1076-2025, Z-1092-2025, Z-1080-2025, Z-1084-2025, Z-1081-2025, Z-1094-2025, Z-1083-2025, Z-1086-2025, Z-1089-2025, Z-1078-2025 and 7 more

Ongoing Class I Medical devices

Recall started
13 January 2025
Reason
Possibility of the patient falling from the table related to the mattress used on the patient table.
Distribution
Worldiwde distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1145-2025, Z-1154-2025, Z-1151-2025, Z-1160-2025, Z-1153-2025, Z-1144-2025, Z-1146-2025, Z-1149-2025, Z-1155-2025, Z-1147-2025, Z-1158-2025, Z-1157-2025 and 8 more

Ongoing Class II Medical devices

Recall started
11 November 2024
Reason
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, Franc
Initiated
Voluntary: Firm initiated
Recall number
Z-0676-2025

Ongoing Class II Medical devices

Recall started
7 October 2024
Reason
Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Isla
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0352-2025, Z-0353-2025

Ongoing Class II Medical devices

Recall started
22 August 2024
Reason
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Distribution
Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3165-2024, Z-3167-2024, Z-3166-2024, Z-3168-2024

Ongoing Class II Medical devices

Recall started
8 August 2024
Reason
A half-threaded bolt was included in the LTE kits instead of the correct full-threaded bolt.
Distribution
Worldwide - US Nationwide distribution in the states of IL and OR. The countries of Hungary, Ireland, Poland, UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-2716-2024

Ongoing Class II Medical devices

Recall started
6 August 2024
Reason
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2953-2024

Ongoing Class II Medical devices

Recall started
15 July 2024
Reason
Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cyprus, Cze
Initiated
Voluntary: Firm initiated
Recall number
Z-2534-2024

Ongoing Class II Medical devices

Recall started
16 May 2024
Reason
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Distribution
Domestic: OK International: India, Spain, U.A.E.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1896-2024, Z-1897-2024, Z-1895-2024

Ongoing Class II Medical devices

Recall started
2 May 2024
Reason
Potential for pressure wave data synchronization from two devices may not be correct.
Distribution
Worldwide distribution - US Nationwide in the states of CA, IN, OR and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, Costa Rica, C¿te D'Ivoire, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hungary, India, Ireland, Italy, Malaysia, Martinique, N
Initiated
Voluntary: Firm initiated
Recall number
Z-1926-2024

Ongoing Class II Medical devices

Recall started
4 April 2024
Reason
Study data is not able to be archived, copied, or exported with the cardiovascular software version.
Distribution
US Nationwide distribution in the states of AK, AL, AZ, CA, CT, FL, GA, IA, IN, KY, MA, MI, MN, NC, NE, NY, OH, OR, TN, TX, VA, and WI. There was government distribution but no military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1252-2025

Ongoing Class II Medical devices

Recall started
6 February 2024
Reason
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Repub
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1178-2024, Z-1176-2024, Z-1177-2024

Ongoing Class II Medical devices

Recall started
2 January 2024
Reason
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1145-2024, Z-1144-2024

Ongoing Class II Medical devices

Recall started
2 January 2024
Reason
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Distribution
Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY; and OUS (International) to countries of: Af
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1152-2024, Z-1151-2024

Ongoing Class II Medical devices

Recall started
2 January 2024
Reason
Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Distribution
Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY; and OUS (International) to countries of: Af
Initiated
Voluntary: Firm initiated
Recall number
Z-1156-2024

Ongoing Class II Medical devices

Recall started
12 December 2023
Reason
System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0835-2024

Ongoing Class II Medical devices

Recall started
8 December 2023
Reason
Loss of connectivity between the FlexArm and the Table due to a software issue.
Distribution
Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0860-2024

Ongoing Class II Medical devices

Recall started
8 December 2023
Reason
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
Distribution
Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0861-2024

Ongoing Class II Medical devices

Recall started
7 December 2023
Reason
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Distribution
US Nationwide distribution.
Recall numbers
Z-2285-2026, Z-2284-2026, Z-2288-2026, Z-2290-2026, Z-2287-2026, Z-2291-2026, Z-2289-2026, Z-2286-2026

Ongoing Class II Medical devices

Recall started
3 November 2023
Reason
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
Distribution
Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Rep
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0559-2024, Z-0558-2024

Ongoing Class II Medical devices

Recall started
15 August 2023
Reason
A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
Distribution
Worldwide distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2587-2023

Ongoing Class II Medical devices

Recall started
20 July 2023
Reason
Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
Distribution
US Nationwide distribution
Initiated
FDA Mandated
Recall number
Z-1583-2025

Ongoing Class II Medical devices

Recall started
19 July 2023
Reason
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall numbers
Z-2402-2023, Z-2403-2023, Z-2404-2023

Ongoing Class II Medical devices

Recall started
30 June 2023
Reason
Loss of availability of the wireless foot switch during procedures.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Chile, China, Croatia, Czech Republic, Denmark, Egypt, Fiji, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kor
Initiated
Voluntary: Firm initiated
Recall number
Z-2485-2023

Ongoing Class II Medical devices

Recall started
12 June 2023
Reason
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Indonesia, Iran, Ireland, Italy, Japan, Jordan, Kenya, South Korea, Libya, Lithuania, Malaysi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2284-2023, Z-2283-2023

Ongoing Class II Medical devices

Recall started
21 March 2023
Reason
Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
Distribution
Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece
Initiated
Voluntary: Firm initiated
Recall number
Z-1647-2023

Ongoing Class II Medical devices

Recall started
20 February 2023
Reason
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1416-2023, Z-1417-2023

Ongoing Class II Medical devices

Recall started
2 December 2021
Reason
Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
Distribution
Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0647-2022, Z-0649-2022, Z-0648-2022

Ongoing Class II Medical devices

Recall started
26 May 2021
Reason
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
Distribution
US Nationwide and Worldwide Distribution
Initiated
FDA Mandated
Recall number
Z-0886-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.