Philips Medical Systems Gmbh, DMC recalls

7 recalls recorded since 2018, most recently on 1 October 2025. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)4
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
1 October 2025
Reason
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
Distribution
Worldwide Distribution: US (Nationwide) Distribution and OUS (International) countries of: Australia, Austria, Bahrain, Belgium, Bosnia-Herz., Canada, Chile, Croatia, Denmark, Estonia, Finland, France, Germany, Ghana, Gibraltar, Guadeloupe, Indonesia, Iraq, Israel, Italy, Jordan, Kuwait, Kyrgyzstan,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0466-2026, Z-0465-2026, Z-0463-2026, Z-0461-2026, Z-0462-2026, Z-0464-2026

Ongoing Class II Medical devices

Recall started
19 May 2021
Reason
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall numbers
Z-1326-2022, Z-1325-2022, Z-1328-2022, Z-1329-2022, Z-1327-2022

Ongoing Class II Medical devices

Recall started
22 July 2020
Reason
Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-2357-2021

Ongoing Class II Medical devices

Recall started
19 December 2019
Terminated
18 June 2020
Reason
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
Distribution
Worldwide distribution  US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, E
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1328-2020, Z-1329-2020

Terminated Class II Medical devices

Recall started
7 November 2018
Terminated
2 September 2020
Reason
kV/mA lockin not functioning as specified which may result in the patient receiving an increased radiation dose.
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, GA, IL, MN, MO, MT, NJ, OH, TX, UT & VA; and to countries of: Australia, Austria, Bangladesh, Belgium, Chile C¿te D'Ivoire, Czech Republic. Egypt, Ethiopia, France, Germany, Ghana, Iran, Italy, Jordan, Latvia, Lebanon, Luxembourg, Malaysi
Initiated
Voluntary: Firm initiated
Recall number
Z-0612-2019

Terminated Class II Medical devices

Recall started
19 September 2018
Terminated
10 August 2023
Reason
Stitching run on SkyPlate detector interrupted intermittently; Stitching with SkyPlate aborts after first image; System software CD contains wrong baseline
Distribution
Worldwide distribution - US Nationwide and countries of Canada Argentina Australia Austria Bangladesh Belgium Bolivia Brazil Chile China Colombia Czech Republic Denmark Egypt Finland France French Guiana Germany Guadeloupe Hong Kong Hungary India Indonesia Iran Israel Ital
Initiated
Voluntary: Firm initiated
Recall number
Z-0239-2019

Terminated Class II Medical devices

Recall started
27 June 2018
Terminated
30 April 2024
Reason
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Distribution
US Nationwide; International to 66 countries
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2846-2018, Z-2858-2018, Z-2843-2018, Z-2863-2018, Z-2867-2018, Z-2842-2018, Z-2862-2018, Z-2854-2018, Z-2851-2018, Z-2861-2018, Z-2848-2018, Z-2844-2018 and 16 more

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.