Andover, MA, United States
Philips Healthcare recalls
7 recalls recorded since 2018, most recently on 2 August 2023. 3 are listed as ongoing. Status and classification are as published by the FDA.
Recalls7
Ongoing (FDA)3
Terminated or completed4
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 2 August 2023
- Reason
- Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.
- Distribution
- US Nationwide Distribution
- Initiated
- FDA Mandated
- Recall number
- Z-2520-2023
- Incisive CT, software version 5.0
Terminated Class II Medical devices
- Recall started
- 2 February 2022
- Terminated
- 5 November 2024
- Reason
- Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.
- Distribution
- US Nationwide distribution.
- Initiated
- FDA Mandated
- Recall number
- Z-0737-2022
- Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x
Ongoing Class II Medical devices
- Recall started
- 24 September 2021
- Reason
- 3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.
- Distribution
- US Nationwide distribution
- Initiated
- FDA Mandated
- Recall number
- Z-0238-2022
- Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)
Ongoing Class II Medical devices
- Recall started
- 22 September 2021
- Reason
- The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control certification statement [21 CFR 1010.2]. Philips has determined that some systems at customer locations did not have these labels applied.
- Distribution
- US Nationwide Distribution
- Initiated
- FDA Mandated
- Recall numbers
- Z-0241-2022, Z-0239-2022, Z-0240-2022
- CombiDiagnost R90 is multi-functional general R/F systems.
- DigitalDiagnost C90
- ProxiDiagnost N90 is multi-functional general R/F systems.
Completed Class II Medical devices
- Recall started
- 26 May 2021
- Reason
- In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
- Distribution
- US Nationwide Distribution
- Initiated
- FDA Mandated
- Recall number
- Z-0734-2022
- Azurion systems with software release R2.0.x
Terminated Class II Medical devices
- Recall started
- 21 December 2018
- Terminated
- 27 October 2020
- Reason
- During recent evaluations of the Philips DigitalDiagnost C50 system, Philips identified a missing label. This notification is intended to provide you with information regarding what the issue is, and relevant affected product, the actions you are required to take and the actions planned by Philips to correct the issue. If you need any or further information or support concerning this issue, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
- Distribution
- Puerto Rico
- Initiated
- FDA Mandated
- Recall number
- Z-2498-2019
- DigitalDiagnost C50, Stationary X-ray System
Terminated Class II Medical devices
- Recall started
- 11 June 2018
- Terminated
- 3 July 2024
- Reason
- The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
- Distribution
- US Nationwide Distribution
- Initiated
- FDA Mandated
- Recall numbers
- Z-2500-2018, Z-2499-2018, Z-2498-2018
- Centron Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, in
- UNIQ Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, inter
- Allura Xper Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.