Philips Electronics North America Corporation recalls

22 recalls recorded since 2018, most recently on 15 November 2018. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls22
Ongoing (FDA)3
Terminated or completed19
Class I (FDA)0

Terminated Class II Medical devices

Recall started
15 November 2018
Terminated
20 September 2024
Reason
There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
Distribution
worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0614-2019

Ongoing Class II Medical devices

Recall started
10 October 2018
Reason
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0592-2019

Terminated Class II Medical devices

Recall started
10 October 2018
Terminated
19 August 2022
Reason
Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.
Distribution
Worldwide Distribution - U.S Nationwide in the states of: GA, MO, IN, FL, NY, HI, CA, LA, WA, IL, NE; Foreign (OUS):Australia, Austria,Belgium, Canada, France, Germany, Hong Kong, Italy, Japan, Kazakhstan, Netherlands, Norway, Spain, Switzerland, Taiwan, United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-0175-2019

Terminated Class II Medical devices

Recall started
16 July 2018
Terminated
18 November 2022
Reason
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
Distribution
Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and countries of: Algeria, Argentina,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3230-2018, Z-3215-2018, Z-3225-2018, Z-3237-2018, Z-3218-2018, Z-3231-2018, Z-3221-2018, Z-3214-2018, Z-3222-2018, Z-3216-2018, Z-3220-2018, Z-3235-2018 and 15 more

Ongoing Class II Medical devices

Recall started
13 July 2018
Reason
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2895-2018, Z-2892-2018, Z-2894-2018, Z-2896-2018, Z-2893-2018

Terminated Class II Medical devices

Recall started
3 May 2018
Terminated
17 October 2019
Reason
The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.
Distribution
Worldwide Distribution to accounts in OH and NC. Foreign distribution to Lebanon.
Initiated
Voluntary: Firm initiated
Recall number
Z-2035-2018

Terminated Class II Medical devices

Recall started
30 April 2018
Terminated
4 October 2019
Reason
Five warning statements are missing from the instructions for use.
Distribution
Worldwide Distribution -- US, to the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and to the territories of Guam and Puerto Rico; a
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2143-2018, Z-2141-2018, Z-2144-2018, Z-2142-2018

Terminated Class II Medical devices

Recall started
16 March 2018
Terminated
10 April 2020
Reason
Potential risk for helium gas inside the MR examination room during a magnet quench
Distribution
Worldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile, China, Colombia, Congo (Democratic Republic
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1438-2018, Z-1430-2018, Z-1436-2018, Z-1435-2018, Z-1440-2018, Z-1431-2018, Z-1434-2018, Z-1441-2018, Z-1437-2018, Z-1432-2018, Z-1433-2018, Z-1442-2018 and 1 more

Terminated Class II Medical devices

Recall started
14 March 2018
Terminated
29 April 2024
Reason
After continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.
Distribution
Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The product was distributed to the following countries: Argentina, Austral
Initiated
Voluntary: Firm initiated
Recall number
Z-1589-2018

Terminated Class II Medical devices

Recall started
14 March 2018
Terminated
19 August 2020
Reason
An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.
Distribution
Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussala
Initiated
FDA Mandated
Recall numbers
Z-1913-2018, Z-1914-2018

Terminated Class III Medical devices

Recall started
7 February 2018
Terminated
7 February 2020
Reason
Update XL¿ device software to version A.03. This includes enhancements to the Operational Checks, event logs, and troubleshooting messages that provide as complete of information as needed to users on device readiness.
Distribution
Foreign Only - China
Initiated
Voluntary: Firm initiated
Recall number
Z-0978-2018

Terminated Class II Medical devices

Recall started
29 January 2018
Terminated
5 May 2020
Reason
Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No. 1:17-cv-11955-DJC, entered October 31, 2017 between United States of America and Philips North America
Distribution
Worldwide Distribution - US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0683-2018, Z-0684-2018

Terminated Class II Medical devices

Recall started
4 January 2018
Terminated
9 October 2019
Reason
There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.
Distribution
Worldwide Distribution - US Nationwide; foreign distribution to Brazil, Germany, Iceland, Japan, Mexico, and (updated 6/25/18) Saudi Arabia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0716-2018

Terminated Class II Medical devices

Recall started
3 January 2018
Terminated
14 November 2019
Reason
Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.
Distribution
Domestic and Foreign
Initiated
Voluntary: Firm initiated
Recall number
Z-0776-2018

Terminated Class II Medical devices

Recall started
20 December 2017
Terminated
16 June 2020
Reason
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
Distribution
Worldwide Distribution - US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0717-2018

Terminated Class II Medical devices

Recall started
11 December 2017
Terminated
29 May 2020
Reason
It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary.
Distribution
Worldwide Distribution - USA (nationwide) Distribution, Internationally to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0704-2018

Terminated Class II Medical devices

Recall started
17 November 2017
Terminated
16 September 2020
Reason
The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.
Distribution
48 Foreign Accounts
Initiated
Voluntary: Firm initiated
Recall number
Z-1315-2018

Terminated Class II Medical devices

Recall started
19 October 2017
Terminated
20 September 2018
Reason
The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
Distribution
Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
Initiated
Voluntary: Firm initiated
Recall number
Z-0330-2018

Terminated Class II Medical devices

Recall started
5 October 2017
Terminated
1 July 2019
Reason
Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0850-2018

Ongoing Class II Medical devices

Recall started
23 August 2017
Reason
Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors
Distribution
Worldwide Distribution and US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0292-2018

Terminated Class II Medical devices

Recall started
17 July 2017
Terminated
12 July 2018
Reason
Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs): " On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave. " On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*. *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+. If the user follows this warning, this problem cannot occur on the XL+.
Distribution
1067 US customers 163 Canada, 917 International
Initiated
Voluntary: Firm initiated
Recall number
Z-0348-2018

Terminated Class II Medical devices

Recall started
14 July 2017
Terminated
5 June 2020
Reason
If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
Distribution
412 units affected.
Initiated
Voluntary: Firm initiated
Recall number
Z-0311-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.