Andover, MA, United States
Philips Electronics North America Corporation recalls
22 recalls recorded since 2018, most recently on 15 November 2018. 3 are listed as ongoing. Status and classification are as published by the FDA.
Recalls22
Ongoing (FDA)3
Terminated or completed19
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 15 November 2018
- Terminated
- 20 September 2024
- Reason
- There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
- Distribution
- worldwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0614-2019
- Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081
Ongoing Class II Medical devices
- Recall started
- 10 October 2018
- Reason
- A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
- Distribution
- US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0592-2019
- IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Terminated Class II Medical devices
- Recall started
- 10 October 2018
- Terminated
- 19 August 2022
- Reason
- Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.
- Distribution
- Worldwide Distribution - U.S Nationwide in the states of: GA, MO, IN, FL, NY, HI, CA, LA, WA, IL, NE; Foreign (OUS):Australia, Austria,Belgium, Canada, France, Germany, Hong Kong, Italy, Japan, Kazakhstan, Netherlands, Norway, Spain, Switzerland, Taiwan, United Kingdom
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0175-2019
- Philips HeartStart FR3 Defibrillator, Model: 861388, 861389 Product Usage: The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA),
Terminated Class II Medical devices
- Recall started
- 16 July 2018
- Terminated
- 18 November 2022
- Reason
- There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and countries of: Algeria, Argentina,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-3230-2018, Z-3215-2018, Z-3225-2018, Z-3237-2018, Z-3218-2018, Z-3231-2018, Z-3221-2018, Z-3214-2018, Z-3222-2018, Z-3216-2018, Z-3220-2018, Z-3235-2018 and 15 more
- AIIuraXperFDlO/10, System Code 722011 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- INTEGRIS CV, System Code 722030 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- Allura Xper FD1O C, System Code 722001 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- Poly C- DMCP-Visub(H3000), System Code 72238 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- INTEGRIS SUITE, System Code 722199 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- AIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- Allura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- INTEGRIS Allura 9 (biplane), System Code 722021 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- and 19 more product lines
Ongoing Class II Medical devices
- Recall started
- 13 July 2018
- Reason
- Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2895-2018, Z-2892-2018, Z-2894-2018, Z-2896-2018, Z-2893-2018
- Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 863077 863085 863086 863317 863287 863288 863289 The SureSigns VM4, VM6, and VM8 patient monitors are for use by health car
- Philips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neonatal patients, and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric
- Philips SureSigns VS4 Vital Signs Monitor, 863283 863286 The SureSigns VS4 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.
- Philips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological pa
- Philips SureSigns VS3 Vital Signs Monitor, 863069 863070 863071 863072 863073 863074 The SureSigns VS3 monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics,
Terminated Class II Medical devices
- Recall started
- 3 May 2018
- Terminated
- 17 October 2019
- Reason
- The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.
- Distribution
- Worldwide Distribution to accounts in OH and NC. Foreign distribution to Lebanon.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2035-2018
- HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable) Product Usage: The HeartStart MRx is for use for the termination of ventric
Terminated Class II Medical devices
- Recall started
- 30 April 2018
- Terminated
- 4 October 2019
- Reason
- Five warning statements are missing from the instructions for use.
- Distribution
- Worldwide Distribution -- US, to the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and to the territories of Guam and Puerto Rico; a
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2143-2018, Z-2141-2018, Z-2144-2018, Z-2142-2018
- IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865352
- IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865350
- IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 867146
- IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865351
Terminated Class II Medical devices
- Recall started
- 16 March 2018
- Terminated
- 10 April 2020
- Reason
- Potential risk for helium gas inside the MR examination room during a magnet quench
- Distribution
- Worldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile, China, Colombia, Congo (Democratic Republic
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1438-2018, Z-1430-2018, Z-1436-2018, Z-1435-2018, Z-1440-2018, Z-1431-2018, Z-1434-2018, Z-1441-2018, Z-1437-2018, Z-1432-2018, Z-1433-2018, Z-1442-2018 and 1 more
- T5-NT, Model 78104 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or
- Multiva systems, 1) Multiva 1.5T 8 R5 Model 781072, 2) Multiva 1.5T 16 R5 Model 781073, 3) Multiva 1.5T 8ch R5 based Model 781074, 4) Multiva 1.5T Model 781076, & 5) Multiva 1.5T 16ch R5 based Model 781078 Product Us
- Conversion SmarthPath to dStream for 1.5T, Model 781260 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the in
- Achieva 1.5T: 1) Nova (Dual), Model 781173, 2) Achieva 1.5T Initial Systems Model 781178, 3) Achieva 1.5T Nova Model 781196, 4) Achieva 1.5T Conversion Model 781283, 5) Achieva 1.5T Model 781296, 6) Achieva 1.5T R5 based
- T10-NT, Model 78107 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,o
- Intera 0.5T Standard, Model 781101 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of t
- Panorama 1.0T, Model 781250 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head
- Multiva systems; 1) Magnets Multiva 16 Model 781487, 2) Magnets 8 Model 781488, 3) Magnet Achieva 1.5T, Model 781492, 4) Magnet HFO, Model 781495, 5) Magnet Multiva 16, Model 781496, 6) Magnet Multiva 8, Model 781497 & 7
- and 5 more product lines
Terminated Class II Medical devices
- Recall started
- 14 March 2018
- Terminated
- 29 April 2024
- Reason
- After continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.
- Distribution
- Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The product was distributed to the following countries: Argentina, Austral
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1589-2018
- Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform
Terminated Class II Medical devices
- Recall started
- 14 March 2018
- Terminated
- 19 August 2020
- Reason
- An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.
- Distribution
- Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussala
- Initiated
- FDA Mandated
- Recall numbers
- Z-1913-2018, Z-1914-2018
- Philips HeartStart FRx AED.
- Philips HeartStart (HS1) Onsite/Home AED
Terminated Class III Medical devices
- Recall started
- 7 February 2018
- Terminated
- 7 February 2020
- Reason
- Update XL¿ device software to version A.03. This includes enhancements to the Operational Checks, event logs, and troubleshooting messages that provide as complete of information as needed to users on device readiness.
- Distribution
- Foreign Only - China
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0978-2018
- HeartStart XL+ Defibrillator/Monitor The HeartStart XL+ is intended for use in a hospital setting by qualified medical personnel trained in the operation of the device and qualified by training in basic life support,
Terminated Class II Medical devices
- Recall started
- 29 January 2018
- Terminated
- 5 May 2020
- Reason
- Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No. 1:17-cv-11955-DJC, entered October 31, 2017 between United States of America and Philips North America
- Distribution
- Worldwide Distribution - US Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0683-2018, Z-0684-2018
- Philips QCPR Meter - used with FR3 Product Usage: The Philips CPR meter is intended for use with the Philips HeartStart FR3 automated external defibrillators (FR3) with PR2.0 or higher software or on HeartStart MRx d
- Philips QCPR Meter - used with MRx Product Usage: The Philips CPR meter is intended for use with the Philips HeartStart FR3 automated external defibrillators (FR3) with PR2.0 or higher software or on HeartStart MRx d
Terminated Class II Medical devices
- Recall started
- 4 January 2018
- Terminated
- 9 October 2019
- Reason
- There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.
- Distribution
- Worldwide Distribution - US Nationwide; foreign distribution to Brazil, Germany, Iceland, Japan, Mexico, and (updated 6/25/18) Saudi Arabia.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0716-2018
- Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular
Terminated Class II Medical devices
- Recall started
- 3 January 2018
- Terminated
- 14 November 2019
- Reason
- Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.
- Distribution
- Domestic and Foreign
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0776-2018
- Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue In
Terminated Class II Medical devices
- Recall started
- 20 December 2017
- Terminated
- 16 June 2020
- Reason
- Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
- Distribution
- Worldwide Distribution - US Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0717-2018
- Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and vascular X-ray imaging applications, including diagnostic,
Terminated Class II Medical devices
- Recall started
- 11 December 2017
- Terminated
- 29 May 2020
- Reason
- It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary.
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution, Internationally to Canada.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0704-2018
- digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imager) As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital
Terminated Class II Medical devices
- Recall started
- 17 November 2017
- Terminated
- 16 September 2020
- Reason
- The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.
- Distribution
- 48 Foreign Accounts
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1315-2018
- IntelliVue X3 Patient Monitor.
Terminated Class II Medical devices
- Recall started
- 19 October 2017
- Terminated
- 20 September 2018
- Reason
- The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
- Distribution
- Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0330-2018
- CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas li
Terminated Class II Medical devices
- Recall started
- 5 October 2017
- Terminated
- 1 July 2019
- Reason
- Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.
- Distribution
- US Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0850-2018
- Philips Network Firewall (Cisco ASA 5506), Model # 867098
Ongoing Class II Medical devices
- Recall started
- 23 August 2017
- Reason
- Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors
- Distribution
- Worldwide Distribution and US Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0292-2018
- IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care pr
Terminated Class II Medical devices
- Recall started
- 17 July 2017
- Terminated
- 12 July 2018
- Reason
- Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs): " On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave. " On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*. *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+. If the user follows this warning, this problem cannot occur on the XL+.
- Distribution
- 1067 US customers 163 Canada, 917 International
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0348-2018
- ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641. Manufactured - September 4, 2002 Octob
Terminated Class II Medical devices
- Recall started
- 14 July 2017
- Terminated
- 5 June 2020
- Reason
- If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
- Distribution
- 412 units affected.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0311-2018
- Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a gri
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.