Percussionaire Corporation recalls

8 recalls recorded since 2018, most recently on 18 August 2026. 7 are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)7
Terminated or completed1
Class I (FDA)6

Ongoing Class II Medical devices

Recall started
18 August 2026
Reason
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Distribution
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, C
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3325-2026, Z-3322-2026, Z-3320-2026, Z-3323-2026, Z-3324-2026, Z-3321-2026, Z-3319-2026

Ongoing Class I Medical devices

Recall started
17 September 2025
Reason
IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.
Distribution
Worldwide distribution. US nationwide, Russia, Japan, Switzerland, Canada, Qatar, Chile, South Africa, United Kingdom, Israel, Turkey, and UAE
Initiated
Voluntary: Firm initiated
Recall number
Z-0031-2026

Ongoing Class I Medical devices

Recall started
23 December 2024
Reason
Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
Distribution
Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.
Initiated
Voluntary: Firm initiated
Recall number
Z-0960-2025

Ongoing Class I Medical devices

Recall started
16 September 2024
Reason
As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.
Distribution
U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire
Initiated
Voluntary: Firm initiated
Recall number
Z-0208-2025

Ongoing Class I Medical devices

Recall started
21 August 2024
Reason
Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.
Distribution
US: Nationwide OUS: Israel Japan Russia Switzerland Canada Turkey
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3208-2024, Z-3207-2024

Ongoing Class I Medical devices

Recall started
12 December 2023
Reason
Due to an assembly error within the breathing circuit, their pressure output of the breathing circuit kit can be affected and have severe impact to the end user
Distribution
US:AL AR CA CT FL GA IL IN KS LA MI MN MO NC NJ OH OR TN TX UT VA OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-0637-2024

Ongoing Class I Medical devices

Recall started
28 March 2023
Reason
Breathing circuit reservoir bag may be assembled at the wrong attachment point, which will result in less than optimal therapy.
Distribution
US: RI, CA, TX
Initiated
Voluntary: Firm initiated
Recall number
Z-0693-2024

Completed Class II Medical devices

Recall started
14 September 2020
Reason
Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to install PAL microprocessor chips and firmware.
Distribution
US: CA, ND, MS, MD
Initiated
Voluntary: Firm initiated
Recall number
Z-0259-2024

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.