Pentax of America Inc recalls

11 recalls recorded since 2018, most recently on 16 July 2025. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls11
Ongoing (FDA)3
Terminated or completed8
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
16 July 2025
Reason
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
Distribution
US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2329-2025

Ongoing Class II Medical devices

Recall started
29 January 2025
Reason
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1238-2025, Z-1237-2025, Z-1236-2025

Ongoing Class II Medical devices

Recall started
14 July 2021
Reason
Updated Reprocessing Instruction For Use (rIFU): Impacted PENTAX Medical Video Upper GI Scopes (EG) and Video Colonoscopes (EC) families
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2449-2021, Z-2448-2021

Terminated Class II Medical devices

Recall started
21 April 2021
Terminated
5 February 2024
Reason
IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal end cap (OE-A63) unexpectedly becoming detached during a procedure and updated to notify users of what immediate actions should be taken in case the event occurs. This can result in unforseen events such as mucosal injury, lacerations, or bleeding of the patient. Detachment of the distal end cap (OE-A63) into the oral cavity of the patient may also result in aspiration
Distribution
US Nationwide distribution in the states of Alabama, California, Louisiana, Maryland, Michigan, Missouri, New Jersey, North Carolina, Pennsylvania, Texas, Washington, D.C.
Initiated
Voluntary: Firm initiated
Recall number
Z-1682-2021

Terminated Class II Medical devices

Recall started
22 September 2020
Terminated
16 May 2022
Reason
There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.
Distribution
US Nationwide and Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0292-2021, Z-0293-2021

Terminated Class II Medical devices

Recall started
3 January 2020
Terminated
26 January 2021
Reason
Distributed in the USA without an approved 510K
Distribution
NC, MA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1268-2020, Z-1267-2020

Terminated Class II Medical devices

Recall started
3 December 2018
Terminated
14 August 2020
Reason
Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.
Distribution
Distribution was made to CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada, Germany, and the Netherlands.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1325-2019, Z-1326-2019, Z-1327-2019

Terminated Class II Medical devices

Recall started
3 December 2018
Terminated
21 August 2020
Reason
The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.
Distribution
Worldwide Distribution: US (nationwide) to states of:: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA;; and countries: Canada, Germany, and Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-0970-2019

Terminated Class II Medical devices

Recall started
30 March 2018
Terminated
28 April 2021
Reason
The catheter alignment arrow printed onto the strain relief is not aligned with the detachment feature of the catheter connector, resulting in the potential for the catheter to be attached to the controller in the detach orientation. In this orientation, the catheter is not securely attached to the controller and may be partially (~1cm) ejected from controller.
Distribution
Distribution US nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-2256-2018

Terminated Class II Medical devices

Recall started
13 February 2018
Terminated
9 October 2020
Reason
The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in order to mitigate the potential risk of infection in flexible endoscopy.
Distribution
USA (nationwide) Distribution to the states of including Puerto Rico : AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and WY, and the District of Columbia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0643-2018

Terminated Class II Medical devices

Recall started
13 July 2017
Terminated
22 July 2019
Reason
The video cytoscopes lack 510(k) premarket notification clearance.
Distribution
Distributed in 8 states: CA, IL, MD, MI, ND, NH, NY, SD, and Guam.
Initiated
Voluntary: Firm initiated
Recall number
Z-0615-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.