Pega Medical Inc. recalls

3 recalls recorded since 2018, most recently on 3 April 2020. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)0
Terminated or completed3
Class I (FDA)0

Terminated Class III Medical devices

Recall started
3 April 2020
Terminated
20 November 2020
Reason
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
Distribution
US Nationwide Distribution including in the states of Florida. No Foreign Consignees listed.
Initiated
Voluntary: Firm initiated
Recall number
Z-2038-2020

Terminated Class II Medical devices

Recall started
2 October 2019
Terminated
26 October 2020
Reason
There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.
Distribution
Nationwide distribution to CA, FL, GA, HI, LA, MA, MO, NC, NY, OK. International distribution to AUSTRALIA, CANADA, DENMARK, GERMANY, ISRAEL, MEXICO, ROMANIA, RUSSIA, UNITED ARAB EMIRATES, UNITED KINGDOM
Initiated
Voluntary: Firm initiated
Recall number
Z-0270-2020

Terminated Class II Medical devices

Recall started
29 May 2019
Terminated
15 January 2021
Reason
Handle could jam with the Driver due to the detachment of an internal Connector Ring.
Distribution
CA, DE, FL, IL, NC, NE, NY and TN
Initiated
Voluntary: Firm initiated
Recall number
Z-2278-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.