Recall Record

San Clemente, CA, United States

Pacific Medical Group Inc. recalls

3 recalls recorded since 2018, most recently on 24 March 2021. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
24 March 2021
Reason
Aftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Distribution
Distribution US nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1375-2021

Ongoing Class II Medical devices

Recall started
19 September 2019
Reason
Fetal transducers distributed in advance of receiving 510(k) and international clearances.
Distribution
U.S.: IN, AZ, IL, HI, AK, AR, CA, FL, GA, IA, ID, LA, MA, NC, NE, NM, NY, OK, PA, TN, TX, UT, VA, WA, WV, NV, KY, CO, ME, ND, OH, SC, WI, AL, OR, DC, DE, MD, MI, MN, MO, MT, MS, NJ, NH. O.U.S. (Foreign): CA, CO, FR, GB, IT, MX, UAE
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0983-2020, Z-0981-2020, Z-0984-2020, Z-0985-2020, Z-0980-2020, Z-0979-2020, Z-0982-2020

Ongoing Class II Medical devices

Recall started
11 September 2019
Reason
Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
Distribution
Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1164-2020, Z-1160-2020, Z-1168-2020, Z-1161-2020, Z-1165-2020, Z-1162-2020, Z-1163-2020, Z-1166-2020, Z-1159-2020, Z-1167-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.