Ottobock Orthopedic Industrie recalls
1 recall recorded since 2018, most recently on 10 January 2019. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 10 January 2019
- Terminated
- 22 June 2021
- Reason
- Some of the installed valve-inserts have a diameter smaller than the specification.
- Distribution
- US Nationwide Distribution
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0828-2019, Z-0826-2019, Z-0827-2019
- 21Y15 MagValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prosthet
- 21Y14 PushValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prost
- 21Y14=S PushValve upper part Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air fro
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.