Vienna, N/A, Austria
Otto Bock Healthcare Product recalls
1 recall recorded since 2018, most recently on 12 September 2017. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 12 September 2017
- Terminated
- 19 January 2021
- Reason
- Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.
- Distribution
- AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0815-2018, Z-0814-2018
- OTTO BOCK Kenevo knee joints Model 3C60=ST
- OTTO BOCK Kenevo knee joints Model 3C60
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.