Recall Record

Foothill Ranch, CA, United States

Ossur Americas recalls

5 recalls recorded since 2018, most recently on 29 August 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)1
Terminated or completed4
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
29 August 2022
Reason
Battery may dislodge from the Power Knee
Distribution
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, VA, VT, WA and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1821-2022

Terminated Class II Medical devices

Recall started
30 May 2022
Terminated
27 December 2024
Reason
Due to receiving complaints related to broken (cracked) prosthetic foot with failures occurring in situations where sudden high impact twisting or bending is applied to the foot.
Distribution
Worldwide distribution including US nationwide (including Puerto Rico).
Initiated
Voluntary: Firm initiated
Recall number
Z-1330-2022

Terminated Class II Medical devices

Recall started
9 November 2021
Terminated
18 December 2023
Reason
Instructions for use were updated to include additional information to facilitate the selection and use of the collar device.
Distribution
US Nationwide distribution in the states of CA, NH, CT, RI, MA, FL, DE, NY, NJ, PA, OH, DC, MD, VA, NC, SC, GA, MS, LA, TN, TX, KY, IN, AZ, MI, WI, MN, IL, IA, MO, NE, MT, ID, CO, NM, WA, OR, NV, UT, KS, AK, VT, AR, OK, AL, SD, WV, PR.
Initiated
Voluntary: Firm initiated
Recall number
Z-0380-2022

Terminated Class II Medical devices

Recall started
1 June 2021
Terminated
22 June 2023
Reason
Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.
Distribution
U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Not provided.
Initiated
Voluntary: Firm initiated
Recall number
Z-1896-2021

Terminated Class II Medical devices

Recall started
21 June 2018
Terminated
4 September 2020
Reason
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
Distribution
U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0997-2019, Z-1003-2019, Z-1000-2019, Z-0999-2019, Z-0998-2019, Z-1002-2019, Z-1001-2019, Z-1004-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.