Oscor Inc. recalls

4 recalls recorded since 2018, most recently on 10 June 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)1
Terminated or completed3
Class I (FDA)2

Ongoing Class I Medical devices

Recall started
10 June 2026
Reason
Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Distribution
Distribution in the US state of MA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2678-2026, Z-2679-2026, Z-2680-2026

Completed Class II Medical devices

Recall started
22 December 2021
Reason
There is potential for separation between the threaded hub and threaded cap at the proximal end of the device.
Distribution
Distribution to TN and PA in USA OUS distribution to Czech Republic
Initiated
Voluntary: Firm initiated
Recall number
Z-0646-2022

Terminated Class II Medical devices

Recall started
10 March 2021
Terminated
8 August 2023
Reason
Potential for open seal on sterile product
Distribution
US Nationwide distribution in the state of New York.
Initiated
Voluntary: Firm initiated
Recall number
Z-1329-2021

Terminated Class I Medical devices

Recall started
26 September 2018
Terminated
4 August 2022
Reason
During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss of connectivity or breakage during movement of the cables causing interruption of the pacing system.
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, HI, IN, KS, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PR, RI, TN, TX, UT, VA, WA. Countries of AUSTRIA, BENIN REPUBLIC, BOLIVIA, BRAZIL, BULGARIA, CANADA, COLOMBIA, COSTA RICA, DOMINICAN REPUBLIC, ECUA
Initiated
Voluntary: Firm initiated
Recall number
Z-0391-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.