Recall Record

Veronaverona, N/A, Italy

Orthofix Srl recalls

4 recalls recorded since 2018, most recently on 24 September 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)1
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
24 September 2024
Reason
Retrospective; Their is the potential that implantable intramedullary lengthening nails may bend at the distal tail during insertion resulting in having to be replaced.
Distribution
US: Unknown, OUS: Unknown
Initiated
Voluntary: Firm initiated
Recall number
Z-1977-2025

Terminated Class II Medical devices

Recall started
27 September 2018
Terminated
1 February 2022
Reason
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Distribution
US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, B
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0378-2020, Z-0383-2020, Z-0390-2020, Z-0387-2020, Z-0379-2020, Z-0375-2020, Z-0403-2020, Z-0385-2020, Z-0374-2020, Z-0391-2020, Z-0398-2020, Z-0400-2020 and 19 more

Terminated Class II Medical devices

Recall started
30 January 2018
Terminated
13 May 2021
Reason
The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.
Distribution
Worldwide Distribution - US Distribution within Texas., and to the countries of : Australia, Belgium, Chile, Denmark, France, Germany, Israel, Italy, Japan, South Africa, Spain, Sweden, Switzerland and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2063-2018

Terminated Class II Medical devices

Recall started
5 July 2017
Terminated
13 January 2023
Reason
The power drill torque limiter may break during use in surgery application.
Distribution
Worldwide Distribution -- US, including the states of CA, FL, & PA; and, the countries of Australia, Germany, Finland, France, Italy, Lithuania, Netherlands, Singapore, South Africa, Sweden, Switzerland, & United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0291-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.