Orthofix, Inc recalls

2 recalls recorded since 2018, most recently on 6 April 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Terminated Class II Medical devices

Recall started
6 April 2018
Terminated
30 June 2021
Reason
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Distribution
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2116-2018, Z-2135-2018, Z-2130-2018, Z-2117-2018, Z-2129-2018, Z-2126-2018, Z-2134-2018, Z-2131-2018, Z-2123-2018, Z-2124-2018, Z-2121-2018, Z-2119-2018 and 8 more

Terminated Class II Medical devices

Recall started
23 February 2018
Terminated
18 June 2019
Reason
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1363-2018, Z-1364-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.