OriGen Biomedical, Inc. recalls

7 recalls recorded since 2018, most recently on 19 December 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)1
Terminated or completed6
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
19 December 2023
Reason
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
Distribution
US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
Initiated
Voluntary: Firm initiated
Recall number
Z-1068-2024

Terminated Class III Medical devices

Recall started
28 January 2021
Terminated
19 April 2023
Reason
It was identified that tissue culture bags were incorrectly labeled as free of phthalates.
Distribution
US: TX, LA, IA, NC, OH, NY, MA, MD, IL, PA, CA, KS, TN, RI, OR, FL, NJ, CT, Puerto Rico OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-1409-2021

Terminated Class II Medical devices

Recall started
24 June 2019
Terminated
14 July 2021
Reason
The firm has become aware that they have distributed product that was expired.
Distribution
US: TX
Initiated
Voluntary: Firm initiated
Recall number
Z-0318-2021

Terminated Class II Medical devices

Recall started
21 August 2018
Terminated
15 July 2021
Reason
These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Distribution
Worldwide Distribution: U.S.(nationwide) to states of: IN, NY, CA, MA, TX, OH, KY, MI, MS, NM, NC, VA, WI, TN, CT, FL, AR, MO, GA, LA, DC, OR, MN, SC, OK, AZ, UT, RI, SD, WA, PA, AL, MD, IL, HI; and countries (OUS) of: Argentina, Austria, Belgium, CANADA, Colombia, EU, Germany, India, Italy, Japan,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0184-2019, Z-0180-2019, Z-0179-2019, Z-0182-2019, Z-0183-2019, Z-0181-2019

Terminated Class II Medical devices

Recall started
21 August 2018
Terminated
5 October 2021
Reason
The catheter failed the endotoxin testing.
Distribution
US Nationwide in the states of NY, AR, MO, NC, GA, LA, CO, AL, MA, CA, MI, TX, OH, MD, FL, TN, SC, OR, AZ, UT, RI, SD, PA, NM, CT. Foreign (OUS): United Arab Emirates, Kingdom of Saudi Arabia, New Zealand, Thailand
Initiated
Voluntary: Firm initiated
Recall number
Z-0102-2019

Terminated Class I Medical devices

Recall started
17 January 2017
Terminated
27 July 2021
Reason
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Distribution
US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Pola
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1232-2019, Z-1231-2019, Z-1230-2019, Z-1228-2019, Z-1229-2019, Z-1227-2019

Terminated Class I Medical devices

Recall started
30 March 2015
Terminated
1 March 2024
Reason
Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage
Distribution
Worldwide Distribution - US Nationwide in the states of: AR, AZ, CA, FL, GA, IL, IN, LA, MA, MI, MO, NC, NY, OH, OR, PA, RI, TX, WA, and WI OUS: Belgium, Canada, Germany, Kingdom of Saudi Arabia, Poland, Spain Sweden, Thailand and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-0269-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.