Austin, TX, United States
OriGen Biomedical, Inc. recalls
7 recalls recorded since 2018, most recently on 19 December 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls7
Ongoing (FDA)1
Terminated or completed6
Class I (FDA)2
Ongoing Class II Medical devices
- Recall started
- 19 December 2023
- Reason
- Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
- Distribution
- US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1068-2024
- Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T
Terminated Class III Medical devices
- Recall started
- 28 January 2021
- Terminated
- 19 April 2023
- Reason
- It was identified that tissue culture bags were incorrectly labeled as free of phthalates.
- Distribution
- US: TX, LA, IA, NC, OH, NY, MA, MD, IL, PA, CA, KS, TN, RI, OR, FL, NJ, CT, Puerto Rico OUS: None
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1409-2021
- Evolve Cell Culture Bags, part numbers: EV1000N and EV3000N The Evolve bag is made of Ethyl Vinyl Olefin (EVO) and contains no BPA or latex. The bag can be supplied with tubing sets that can include, but are not lim
Terminated Class II Medical devices
- Recall started
- 24 June 2019
- Terminated
- 14 July 2021
- Reason
- The firm has become aware that they have distributed product that was expired.
- Distribution
- US: TX
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0318-2021
- 3/8" FloPump, 32mL, (Sterile), REF 6400S, UDI: (01) 081432102013 1, Sterile EO, Rx Only
Terminated Class II Medical devices
- Recall started
- 21 August 2018
- Terminated
- 15 July 2021
- Reason
- These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
- Distribution
- Worldwide Distribution: U.S.(nationwide) to states of: IN, NY, CA, MA, TX, OH, KY, MI, MS, NM, NC, VA, WI, TN, CT, FL, AR, MO, GA, LA, DC, OR, MN, SC, OK, AZ, UT, RI, SD, WA, PA, AL, MD, IL, HI; and countries (OUS) of: Argentina, Austria, Belgium, CANADA, Colombia, EU, Germany, India, Italy, Japan,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0184-2019, Z-0180-2019, Z-0179-2019, Z-0182-2019, Z-0183-2019, Z-0181-2019
- Reinforced Dual Lumen Cannula, 32Fr., REF: VV32F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the int
- Reinforced Dual Lumen Cannula, 16Fr., REF: VV16F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the inte
- Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO; and ECMO Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a sing
- Reinforced Dual Lumen Cannula, 23Fr., REF: VV23F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the inte
- ECMO Reinforced Dual Lumen Cannula, 28Fr., REF: VV28F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the
- Reinforced Dual Lumen Cannula, 19Fr., REF: VV19F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the int
Terminated Class II Medical devices
- Recall started
- 21 August 2018
- Terminated
- 5 October 2021
- Reason
- The catheter failed the endotoxin testing.
- Distribution
- US Nationwide in the states of NY, AR, MO, NC, GA, LA, CO, AL, MA, CA, MI, TX, OH, MD, FL, TN, SC, OR, AZ, UT, RI, SD, PA, NM, CT. Foreign (OUS): United Arab Emirates, Kingdom of Saudi Arabia, New Zealand, Thailand
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0102-2019
- Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO; and ECMO Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO
Terminated Class I Medical devices
- Recall started
- 17 January 2017
- Terminated
- 27 July 2021
- Reason
- Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
- Distribution
- US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Pola
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1232-2019, Z-1231-2019, Z-1230-2019, Z-1228-2019, Z-1229-2019, Z-1227-2019
- ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV28F
- ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV19F
- ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F
- ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV23F
- ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV32F
- ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV16F
Terminated Class I Medical devices
- Recall started
- 30 March 2015
- Terminated
- 1 March 2024
- Reason
- Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AR, AZ, CA, FL, GA, IL, IN, LA, MA, MI, MO, NC, NY, OH, OR, PA, RI, TX, WA, and WI OUS: Belgium, Canada, Germany, Kingdom of Saudi Arabia, Poland, Spain Sweden, Thailand and United Kingdom
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0269-2019
- REF VV19F Reinforced Dual Lumen Cannula, 19Fr., Sterile EO, DISPOSABLE: DO NOT RE-USE CE 0473 Product Usage: The OriGen RDLC catheter is indicated for use as a single cannula for both venous drainage and arterial re
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.