Recall Record

Durham, NC, United States

Organon Teknika Inc recalls

1 recall recorded since 2018, most recently on 11 January 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
11 January 2019
Terminated
6 November 2020
Reason
This notice has been initiated due to potential for compromised test card Top Seal Integrity which could: - yield false resistance for antibiotics on the AST panel - cause a false negative ESBL Test - result in a false positive urea (URE) reaction on ID cards
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0962-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.