Recall Record

Cranberry Township, PA, United States

Omnicell, Inc. recalls

3 recalls recorded since 2018, most recently on 25 March 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
25 March 2026
Reason
Potential for mislabeled syringe produced by the i.v.STATION device.
Distribution
US Distribution in Alabama, Maryland and Pennsylvania.
Initiated
Voluntary: Firm initiated
Recall number
Z-2047-2026

Ongoing Class II Medical devices

Recall started
20 February 2025
Reason
IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.
Distribution
Distributed in the US to Alabama
Initiated
Voluntary: Firm initiated
Recall number
Z-1532-2025

Ongoing Class II Medical devices

Recall started
27 March 2023
Reason
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
Distribution
AL, NY
Initiated
Voluntary: Firm initiated
Recall number
Z-1617-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.