Recall Record

Sanford, FL, United States

Omega Medical Imaging, Inc. recalls

2 recalls recorded since 2018, most recently on 8 September 2023. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)2
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 September 2023
Reason
An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.
Distribution
Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma
Initiated
Voluntary: Firm initiated
Recall number
Z-0160-2024

Ongoing Class II Medical devices

Recall started
28 February 2023
Reason
An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.
Distribution
Worldwide distribution - US Nationwide and the country of Pakistan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0156-2024

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.