Wetzlar, N/A, Germany

Oculus Optikgeraete GMBH recalls

3 recalls recorded since 2018, most recently on 8 August 2023. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)2
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 August 2023
Reason
Their is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
Distribution
US Nationwide distribution in the states of CA, FL, MA, MN, NY, NC, OH, PA, TX, VA, WA. OUS: Not Provided
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2483-2023, Z-2484-2023, Z-2482-2023

Ongoing Class II Medical devices

Recall started
8 July 2022
Reason
Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs
Distribution
U.S.: AL, AR, AZ, CAQ, CO, CT, FL, GA, HI, IA, ID, IL, KS, KY, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WY O.U.S.: N/A
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1673-2022, Z-1672-2022, Z-1674-2022

Terminated Class II Medical devices

Recall started
21 November 2017
Terminated
22 March 2018
Reason
The device software versions have an anomaly which may produce an erroneous marking for the quality specification value
Distribution
US Distribution to the states of : CA and FL. There were no foreign/military/government accounts.
Initiated
Voluntary: Firm initiated
Recall number
Z-0635-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.