O-Two Medical Technologies, Inc. recalls

2 recalls recorded since 2018, most recently on 12 February 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)1

Terminated Class I Medical devices

Recall started
12 February 2019
Terminated
29 June 2023
Reason
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Distribution
The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0978-2019, Z-0977-2019, Z-0976-2019

Terminated Class II Medical devices

Recall started
14 May 2018
Terminated
3 June 2020
Reason
Higher levels of nitrous oxide delivered than specification
Distribution
AK, CA, FL
Initiated
Voluntary: Firm initiated
Recall number
Z-2885-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.