Recall Record

Hasselager, N/A, Denmark

NRT X-RAY A/S recalls

3 recalls recorded since 2018, most recently on 17 April 2024. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
17 April 2024
Reason
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
Distribution
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1784-2024

Ongoing Class II Medical devices

Recall started
27 April 2023
Reason
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Distribution
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1669-2023, Z-1668-2023

Ongoing Class II Medical devices

Recall started
21 January 2022
Reason
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.
Distribution
USA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic, Germany, Denmark, Netherlands, RSA, Sweden
Initiated
Voluntary: Firm initiated
Recall number
Z-0526-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.