NOXBOX LTD recalls

6 recalls recorded since 2018, most recently on 14 September 2026. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)5
Terminated or completed1
Class I (FDA)3

Ongoing Class II Medical devices

Recall started
14 September 2026
Reason
Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
Distribution
US Nationwide distribution in the state of Georgia.
Initiated
Voluntary: Firm initiated
Recall number
Z-3306-2026

Ongoing Class II Medical devices

Recall started
19 August 2026
Reason
Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
Distribution
US Nationwide distribution in the state of Georgia.
Initiated
Voluntary: Firm initiated
Recall number
Z-3165-2026

Ongoing Class I Medical devices

Recall started
9 September 2025
Reason
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
Distribution
Domestic: TN
Initiated
Voluntary: Firm initiated
Recall number
Z-0028-2026

Ongoing Class I Medical devices

Recall started
9 September 2025
Reason
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Distribution
Domestic: TN;
Initiated
Voluntary: Firm initiated
Recall number
Z-0029-2026

Ongoing Class II Medical devices

Recall started
19 July 2023
Reason
Replacement of the internal Sample Pump due to high failure rates.
Distribution
US Nationwide distribution in the state of TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-2544-2025

Terminated Class I Medical devices

Recall started
3 May 2023
Terminated
31 October 2024
Reason
Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.
Distribution
Worldwide distribution - US Nationwide and the countries of Bahrain, Bolivia, Brazil, Ecuador, Finland, France, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Panama, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Turkey, UAE, Unite
Initiated
Voluntary: Firm initiated
Recall number
Z-1667-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.