NordicNeuroLab AS recalls

8 recalls recorded since 2018, most recently on 16 June 2022. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)0
Terminated or completed8
Class I (FDA)0

Terminated Class II Medical devices

Recall started
16 June 2022
Terminated
4 September 2024
Reason
When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1462-2022

Terminated Class II Medical devices

Recall started
20 August 2018
Terminated
16 September 2020
Reason
Complaints of this version of the power supply unit (LPS) (1.0 and 1.1) being dragged into the magnet bore (MR Scanner).
Distribution
worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-3023-2018

Terminated Class II Medical devices

Recall started
28 October 2017
Terminated
31 March 2021
Reason
Report was of an interrupted MR scan after patient complained about acoustic distortion, and heat felt form the headphones.
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Russia, Israel. Norway, South Korea, China, France, South Africa, Japan, Poland, Chile, Switzerland. Australia, Belgium, Lithuania, United Kingdom and Bulgaria
Initiated
Voluntary: Firm initiated
Recall number
Z-2103-2018

Terminated Class II Medical devices

Recall started
3 July 2015
Terminated
16 September 2020
Reason
The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the coregistration.
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia , Belgium, Brazil, China, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway and Poland.
Initiated
Voluntary: Firm initiated
Recall number
Z-2061-2018

Terminated Class II Medical devices

Recall started
15 December 2014
Terminated
24 March 2021
Reason
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : SA, Russia, South Korea, UK, Japan, Australia, Sweden, Norway, Greece
Initiated
Voluntary: Firm initiated
Recall number
Z-2042-2018

Terminated Class II Medical devices

Recall started
17 September 2014
Terminated
24 March 2021
Reason
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Distribution
USA, Sweden, Spain, Russia, Hungary, India, Norway, Iceland, Italy, South Korea, United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-2050-2018

Terminated Class II Medical devices

Recall started
12 September 2014
Terminated
6 October 2020
Reason
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Distribution
Worldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2044-2018

Terminated Class II Medical devices

Recall started
21 December 2012
Terminated
16 September 2020
Reason
There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visualized in the 3D viewer. Then a new images series B is visualized in the 3D viewer. If image series A and B have the same geometrical resolution (pixel sizes and slice distances), the coregistration matrix taking A to B will not be applied to the fiber group, and therefore the fiber group will in general not be positioned correctly on B in the 3D viewer. Furthermore, if the fiber group is exported as a new image series the fibers could be misplaced, and similarly if the fibers are visualized in the MPR on B, the positioning will not be correct. The misalignment will be equal to the rotations/translations necessary to align A with B.
Distribution
Worldwide Distribution - US Nationwide - Saudi Arabia, Brazil, Sweden, Spain, Russia, Belgium, Norway, South Korea, Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-2045-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.