Nikon Metrology recalls
1 recall recorded since 2018, most recently on 23 December 2019. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 23 December 2019
- Terminated
- 13 June 2024
- Reason
- The manufacturer identified a risk that their products might fail to comply with the cabinet x-ray system performance standard limit on x-ray emissions [21 CFR 1020.40(c)(1)] due to an insufficient safety margin for the means of attachment of a piece of interior x-ray shielding. Double sided industrial tape was used for attachment - there have been only 2 instances of detachment and although the piece of shielding detached it remained mostly in place supported by wires that pass through that piece of shielding. The repair is accomplished by adding a bracket to fasten the piece of shielding securely to the inside of the x-ray system. Mfr reports that no systems were found to have exceeded the emission limit when checked prior to repair.
- Distribution
- US Nationwide Distribution
- Initiated
- FDA Mandated
- Recall number
- Z-1727-2020
- Nikon Metrology XTH cabinet x-ray Systems which includes variations STH225, XTH225ST, and MCT225.