Nextremity Solutions recalls

7 recalls recorded since 2018, most recently on 10 February 2022. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)0
Terminated or completed7
Class I (FDA)0

Terminated Class II Medical devices

Recall started
10 February 2022
Terminated
10 October 2024
Reason
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Distribution
Distribution US Nationwide, Japan, and the Netherlands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0913-2022, Z-0914-2022, Z-0912-2022, Z-0910-2022, Z-0911-2022

Terminated Class II Medical devices

Recall started
13 January 2022
Terminated
24 May 2024
Reason
During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision
Distribution
US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0739-2022, Z-0740-2022

Terminated Class II Medical devices

Recall started
26 March 2021
Terminated
10 November 2021
Reason
Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended
Distribution
AL, AK, CA, IA,IL, IN, KY ,MD. MI, NJ, NY, OR, PA, SC, TX, VA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1500-2021, Z-1501-2021

Terminated Class II Medical devices

Recall started
12 February 2020
Terminated
17 December 2020
Reason
Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from the proximal Nextra implant
Distribution
International distribution in the countries of Spain, United Kingdom, Germany, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1458-2020

Terminated Class II Medical devices

Recall started
16 December 2019
Terminated
8 September 2020
Reason
The product contains a different plate than indicated by the package labeling. The product is labeled as a right plate but contains a left plate.
Distribution
Domestic distribution to AL, CA, FL, GA, ID, IL, IA, KY, LA, ME, MA, MI, MN, MS, NE, NJ, NY, OH ,OK, OR, TX, and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1127-2020

Terminated Class II Medical devices

Recall started
11 November 2019
Terminated
10 September 2020
Reason
Components 2mm shorter than indicated on the package label
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0778-2020, Z-0776-2020, Z-0777-2020

Terminated Class II Medical devices

Recall started
9 October 2019
Terminated
24 June 2020
Reason
Product labeled as STRM-NL-3550ST, NL LP Screw 3.5x50mm ST, lot M1151 may contain the incorrect screw type and length
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0534-2020