Neusoft Medical Systems Co., Ltd. recalls

1 recall recorded since 2018, most recently on 22 October 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
22 October 2018
Terminated
25 August 2022
Reason
Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
Distribution
US Nationwide Distribution.
Initiated
FDA Mandated
Recall numbers
Z-1279-2019, Z-1281-2019, Z-1280-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.