NeuroLogica Corporation recalls

10 recalls recorded since 2018, most recently on 7 October 2025. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls10
Ongoing (FDA)5
Terminated or completed5
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
7 October 2025
Reason
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
Distribution
Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Ara
Initiated
Voluntary: Firm initiated
Recall number
Z-0480-2026

Ongoing Class II Medical devices

Recall started
8 July 2024
Reason
Potential for the anti-fall system of the device arm to fail.
Distribution
US Nationwide distribution in the states of AR, CT, GA, IN, KS, MA, MD, MO, MS, NC, ND, NJ, NV, OH, PA, VA, WV & DC.
Initiated
Voluntary: Firm initiated
Recall number
Z-2715-2024

Ongoing Class II Medical devices

Recall started
28 December 2023
Reason
Mobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm.
Distribution
US Nationwide Distribution including states of: AL, AR, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SD, TN, TX, VA, WI & WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1056-2024

Ongoing Class II Medical devices

Recall started
6 June 2023
Reason
1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-2068-2023

Ongoing Class II Medical devices

Recall started
8 March 2023
Reason
Issue related to the operation of the arm latch within the column that supports the arm, potential bodily harm due to a fall of the arm
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1414-2023

Terminated Class II Medical devices

Recall started
15 July 2020
Terminated
13 July 2021
Reason
There are potential issues related to the operational state of the scanner, specifically for transport and inadvertent system shutdowns. There could be a possibility of potential delay in patient treatment if proper use instructions are not followed.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, India, Portugal, Singapore, Slovak Republic, Thailand, Turkey, UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-2772-2020

Terminated Class II Medical devices

Recall started
25 July 2019
Terminated
11 September 2020
Reason
There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2276-2019, Z-2275-2019

Terminated Class II Medical devices

Recall started
6 May 2019
Terminated
10 June 2020
Reason
Potential issue related to the operation of the safety latch within the column that supports the arm due to a possibility of potential bodily harm due to a fall of the arm in an unlikely event that the column wire breaks.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-1691-2019

Terminated Class II Medical devices

Recall started
12 January 2018
Terminated
14 August 2018
Reason
While transporting the portable machine with the retrofit equipment in a backwards motion, there is potential for foot injury from the underside of the detector bin of the retrofit unit.
Distribution
US Distribution . Only one consignee was shipped 16 units of the affected device.
Initiated
Voluntary: Firm initiated
Recall number
Z-1391-2018

Terminated Class II Medical devices

Recall started
22 June 2017
Terminated
5 October 2018
Reason
An image was obtained with over exposure during a Thorax examination using the AEC function.
Distribution
Nationwide US
Initiated
Voluntary: Firm initiated
Recall number
Z-0726-2018