Bozeman, MT, United States
Neuralynx Inc recalls
1 recall recorded since 2018, most recently on 14 March 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 14 March 2022
- Reason
- Atlas Stim Headbox (ASHB) was delivered to US customers under a premise that the device had an Investigational Device Exemption (IDE) and was for research use only, but the manufacturer does not have an approved IDE by the FDA. The product has not undergone electrical safety testing and has not received regulatory clearance.
- Distribution
- U.S. Nationwide distribution in the states of AL, CA, IL, MN, NC, RI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1091-2022
- ATLAS Stim Headbox-NK For use only with the ATLAS Neurophysiology System. CAUTION--INVESTIGATIONAL DEVICE Limited by Federal (or United States) law to investigational use. Only for use in Non-Significant Risk studies und