Recall Record

Denver, CO, United States

Navajo Manufacturing Company recalls

1 recall recorded since 2018, most recently on 18 February 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
18 February 2026
Reason
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Distribution
US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1588-2026

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.