Recall Record

Carlsbad, CA, United States

Nalu Medical, Inc. recalls

1 recall recorded since 2018, most recently on 3 June 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
3 June 2024
Reason
Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.
Distribution
US Nationwide distribution in the states of MO, IL, CA, TX, TN, OK, AZ, AR.
Initiated
Voluntary: Firm initiated
Recall number
Z-2257-2024