Moximed, Inc. recalls

1 recall recorded since 2018, most recently on 2 December 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
2 December 2024
Reason
Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.
Distribution
US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0762-2025