Moller Vital recalls
1 recall recorded since 2018, most recently on 4 February 2019. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 4 February 2019
- Reason
- The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.
- Distribution
- Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1352-2019, Z-1351-2019
- Molift 2-Point sling bar, an accessory for Molift Mover 180/Air. Item nos. 1830001 1830002 1830002C 1830003
- Molift 2-Point sling bar, an accessory for Molift Mover 205. Item nos. 1430016 1430017 1430021
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.