Mobius Imaging, LLC recalls

5 recalls recorded since 2018, most recently on 4 December 2025. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)2
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
4 December 2025
Reason
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
Distribution
US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1040-2026, Z-1041-2026

Ongoing Class II Medical devices

Recall started
7 December 2022
Reason
Stryker received one report of an injury to an untrained user that occurred while the untrained user was moving an Airo TruCT unit in reverse. The firm will increase visibility of instructions for users to consider when transporting the unit in reverse by providing labels to be affixed onto the AIRO.
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Chinese Taipei, Denmark, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Malaysia, Mexico, Morocco, Nepal, Nethe
Initiated
Voluntary: Firm initiated
Recall number
Z-0957-2023

Terminated Class II Medical devices

Recall started
18 November 2022
Terminated
26 February 2025
Reason
Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.
Distribution
Domestic distribution to consignees in Colorado, Tennessee, and Utah.
Initiated
Voluntary: Firm initiated
Recall number
Z-0344-2023

Terminated Class II Medical devices

Recall started
1 August 2018
Terminated
13 December 2019
Reason
The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed.
Distribution
US and foreign distribution.
Initiated
FDA Mandated
Recall number
Z-2991-2018

Terminated Class II Medical devices

Recall started
26 September 2017
Terminated
16 August 2018
Reason
If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, NC, NY, OH, OR, PA, RI, TN, TX, UT, VA, WA, WV and the countries of Argentina, Australia, Austria, Belgium, Canada, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Lithuania, Ne
Initiated
Voluntary: Firm initiated
Recall number
Z-0715-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.