mo-Vis BVBA recalls

2 recalls recorded since 2018, most recently on 10 July 2025. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)2
Terminated or completed0
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
10 July 2025
Reason
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Distribution
US Nationwide distribution in the state of TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2456-2025, Z-2454-2025, Z-2459-2025, Z-2458-2025, Z-2457-2025, Z-2455-2025

Ongoing Class II Medical devices

Recall started
28 November 2023
Reason
In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.
Distribution
US States: TN, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0713-2024